start › cdms › k100 › run_diary › 20210603

2021/06/03 (Thursday)

Previous Day

  • Lab in-personnel: Matt, Scott
Status Run 74 in progress. Warmed up after 3rd appearance of IVC leak at 4K. Cryostat Vacuum Jacket pumped out.
Began tower installation yesterday
1455 Check fridge pressures: LHe trap/Sondenser/Still = 1.6/<0.1/0.2 mbar. Pump out briefly with Scroll pump
1500 Check contents of warm LN2 Trap 1: 4.4 mbar. Pump out with Scroll pump.
1505 Check contents of warm LN2 Trap 2: 1.8 mbar. Pump out with Scroll pump.
1535 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.6%
1540 Increase V6 → 12.7% to speed up flow
1625 Stop cleaning through Trap 1 after 50 min. G2 flow pressure was 585-705 mbar. Pump mixture to dump.
Removed 9.8 mbar.
Note: at the end there was no flow, so this cleaning probably only had flow for about 10 or 15 minutes.

Next Day