| 1150 | Perform mixture throughput test at MC = 300 K using above procedure with 3He pump on. Let G1 rise to 560 mbar before Step 3, then it falls to 42 mbar and starts to rise. Throughput = 6.5 mbar/min. |
| 1155 | As on 2021 Apr 15, we will flow mixture through the fridge during LN2 cooldown. Let mixture from dump through cold LN2 Trap 1 to fridge, both sides. Then isolate the sides and circulate with V6 open partially and mix open to dump. V9, V13a, V12a+speedi, V1, Condenser speedi, Still gate valves open, 3He pump on. Open V6 to 12.8%. |
| 1205 | P1(Condenser) → 26 mbar. Strange, this is significantly higher than in the past. I started V6 at 12.6% and ticked it up a bit to 12.8%. I'll leave it there. |
| 1210 | Try closing V12a speedivalve. G1 falls as it should, showing we have throughput. Falls at 11.0 mbar/min . Open V12a speedivalve again. |
| 1215 | Turbo on IVC = 8.6 nbar. With Turbo valved out for 1 minute IVC = 160 nbar. |
| 1220 | Let 5.4 mbar He exchange gas into IVC. |
| 1230 | Begin LN2 transfer |
| 1230 | MC(Cernox) = 293.0 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 682/740/off-scale mbar, Level resistor = 0.987 kΩ. DIB2 qI FB to Chassis = 0.11 nS. |
| 1235 | LN2 Dewar = 8.0 PSI, pretty low, transfer seems slow. Let's pressurize the Dewar with N2 gas. Adjust regulator on N2 cylinder to 11 PSI. |
| 1240 | Opened LN2 Dewar Vent valve enough so that it was making a fairly loud rattling noise. Gauge on Dewar has not increased above 8 PSI, but transfer seems quite fast, significant breeze coming out of the MB exhaust. So close the Vent valve enough so that the noise is mostly gone. |
| 1245 | MC(Cernox) = 290.2 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1 (circulation) = 682/738 mbar, Level resistor = 1.052 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1300 | MC(Cernox) = 273.0 K, Sorb = >249 K (max), IVC = 4.0 mbar, G2/G1 (circulation) = 681/736 mbar, Level resistor = 1.099 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1300 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 14.6 mbar/min. Open V12a speedivalve again. |
| 1315 | MC(Cernox) = 237.7 K, Sorb = >249 K (max), IVC = 3.4 mbar, G2/G1 (circulation) = 678/733 mbar, Level resistor = 1.156 kΩ. DIB2 qI FB to Chassis = 0.10 nS. |
| 1315 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 21.7 mbar/min. Open V12a speedivalve again. |
| 1330 | MC(Cernox) = 197.5 K, Sorb = >249 K (max), IVC = 2.5 mbar, G2/G1 (circulation) = 675/729 mbar, Level resistor = 1.236 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1330 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 23.5 mbar/min. Open V12a speedivalve again. |
| 1345 | MC(Cernox) = 168.0 K, Sorb = >249 K (max), IVC = 2.2 mbar, G2/G1 (circulation) = 671/725 mbar, Level resistor = 1.282 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1345 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 30.8 mbar/min. Open V12a speedivalve again. |
| 1400 | MC(Cernox) = 148.4 K, Sorb = 206.7 K, IVC = 2.2 mbar, G2/G1 (circulation) = 665/719 mbar, Level resistor = 1.305 kΩ. DIB2 qI FB to Chassis = 0.09 nS. |
| 1400 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 37.7 mbar/min. Open V12a speedivalve again. |
| 1415 | MC(Cernox) = 132.7 K, Sorb = 171.7 K, IVC = 1.9 mbar, G2/G1 (circulation) = 659/712 mbar, Level resistor = 1.318 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1415 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 42.6 mbar/min. Open V12a speedivalve again. |
| 1430 | MC(Cernox) = 121.1 K, Sorb = 150.3 K, IVC = 1.9 mbar, G2/G1 (circulation) = 651/704 mbar, Level resistor = 1.326 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1430 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 45.4 mbar/min. Open V12a speedivalve again. |
| | Getting close to where we were when we stopped last time. This transfer is faster. By the Sorb and the Level Resistor, we're already further along than when we stopped last time. But the MC temperature is not quite as low. We could maybe err a little later since it will be sitting longer till tomorrow morning. |
| 1445 | MC(Cernox) = 111.4 K, Sorb = 137.5 K, IVC = 1.9 mbar, G2/G1 (circulation) = 642/694 mbar, Level resistor = 1.329 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1445 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 44.8 mbar/min. Open V12a speedivalve again. |
| 1455 | MC(Cernox) = 106.9 K, Sorb = 133.2 K, IVC = 1.9 mbar, G2/G1 (circulation) = 638/689 mbar, Level resistor = 1.330 kΩ. DIB2 qI FB to Chassis = 0.07 nS. |
| 1455 | Stop LN2 transfer. Close both LN2 Dewar valves and N2 gas supply valve on wall. Pull out Level Resistor and put stopper in port. Put ping-pong ball on MB exhaust port. |
| 1500 | Close V6 and turn off 3He pump. |
| 1505 | Matt exits lab |