| 1130 | Perform mixture throughput test at MC = 300 K using above procedure with 3He pump on. Let G1 rise to 510 mbar before Step 3, then it falls to 38 mbar and starts to rise. Throughput = 5.3 mbar/min. |
| 1140 | Check pressure in 1K Pot: 24 mbar on P2, 0.07 atm on mechanical gauge. Pump out with Scroll Pump |
| 1150 | With 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump. Once P2 < 2 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume to < 1 mbar on P2. |
| 1155 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar. |
| 1200 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar. |
| 1305 | Pressurize main bath and 1K Pot (both sides) with helium to 311 mbar +1atm on G3, then close 1K Pot speedivalve and needle valve. |
| 1305 | As on 2021 Apr 6, we will flow mixture through the fridge during LN2 cooldown. Let mixture from dump through cold LN2 Trap 1 to fridge, both sides. Then isolate the sides and circulate with V6 open partially and mix open to dump. V9, V13a, V12a+speedi, V1, Condenser speedi, Still gate valves open, 3He pump on. Open V6 to 12.6%. |
| 1310 | P1(Condenser) → 0.32 mbar. |
| 1310 | Leak Detector on IVC with Turbo valved out for 1 minute: 31 pbarL/s |
| 1315 | Let 5.4 mbar He exchange gas into IVC. |
| 1315 | Try closing V12a speedivalve. G1 falls as it should, showing we have throughput. Falls at 11.6 mbar/min . Open V12a speedivalve again. |
| 1320 | MC(Cernox) = 296.3 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 672/744/0.32 mbar, Level resistor = 0.982 kΩ. DIB2 qI FB to Chassis = 0.11 nS. |
| 1330 | Begin LN2 transfer. Transfer seems a little weak, we may want to pressurize the Dewar. |
| 1330 | Pressurize LN2 Dewar through the vent port with N2. Dewar gauge = 10.5 PSI, not much different after applying pressure. We could increase it with the cylinder regulator but this will probably be good enough. |
| 1355 | MC(Cernox) = 285.9 K, Sorb = >249 K (max), IVC = 4.0 mbar, G2/G1/P1 (circulation) = 674/744/2.2 mbar, Level resistor = 1.001 kΩ. DIB2 qI FB to Chassis = 0.12 nS. |
| 1410 | MC(Cernox) = 263.1 K, Sorb = >249 K (max), IVC = 3.4 mbar, G2/G1/P1 (circulation) = 675/744/5.0 mbar, Level resistor = 1.023 kΩ. DIB2 qI FB to Chassis = 0.09 nS. |
| 1410 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 14.6 mbar/min. Open V12a speedivalve again. |
| 1425 | MC(Cernox) = 224.2 K, Sorb = >249 K (max), IVC = 2.5 mbar, G2/G1/P1 (circulation) = 674/742/26 mbar, Level resistor = 1.086 kΩ. DIB2 qI FB to Chassis = 0.11 nS. |
| 1425 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 18.6 mbar/min. Open V12a speedivalve again. |
| 1430 | Matt exits lab |
| 1458 | Derek enters lab. |
| 1500 | MC(Cernox) = 153.3 K, Sorb = >249 K (max), IVC = 2.2 mbar, G2/G1/P1 (circulation) = 665/734/1000 mbar, Level resistor = 1.242 kΩ. DIB2 qI FB to Chassis = 0.11 nS. |
| 1504 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 32.8 mbar/min. Open V12a speedivalve again. |
| 1515 | MC(Cernox) = 138.4 K, Sorb = 220.5 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 660/728/1000 mbar, Level resistor = 1.274 kΩ. DIB2 qI FB to Chassis = 0.09 nS. |
| 1530 | MC(Cernox) = 124.5 K, Sorb = 180.5 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 653/720/1000 mbar, Level resistor = 1.299 kΩ, Level resistor (Red MM) = 1.303 . DIB2 qI FB to Chassis = 0.09 nS. |
| 1533 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 46.2 mbar/min. Open V12a speedivalve again. |
| 1530 | MC(Cernox) = 114.9 K, Sorb = 157.3 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 646/712/1000 mbar, Level resistor = 1.313 kΩ, Level resistor (Red MM) = 1.317 . DIB2 qI FB to Chassis = 0.10 nS. |
| 1556 | MC(Cernox) = 108.8 K, Sorb = 145.1 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 640/705/1000 mbar, Level resistor = 1.320 kΩ. DIB2 qI FB to Chassis = 0.09 nS. |
| 1558 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 54.4 mbar/min. Open V12a speedivalve again. |
| 1601 | Stop LN2 transfer. Close both LN2 Dewar valves and N2 gas supply valve on wall. Pull out Level Resistor and put stopper in port. Put ping-pong ball on MB exhaust port. |
| 1606 | Close V6 and turn off 3He pump. |
| 1610 | Derek leaves lab. |