2021/04/22 (Thursday)

Previous Day

  • Lab in-person: Zach
  • Virtual partner: Matt
Status Run 74 in progress. Warming to fix IVC leak. Helium gas in cryostat vacuum jacket. Mixture in dump. N2 in IVC.
0830 MC(Cernox) = 263.6 K, rising at 1.2 K/hour. Suggests tau_warming of 26.5 hours, predicted to reach dew point tonight at 19:12.
1600 MC(Cernox) = 271.0 K, rising at 0.9 K/hour. Suggests tau_warming of 27.1 hours, predicted to reach dew point tonight at 19:44.
1830 Zach arrives at lab.
1835 Mechanical gauge on IVC = -12 InHg.
1850 Switched line to CVJ, begin pumping with scroll then turbo.
1900 CVJ = 0.25 mbar, pump out with scroll pump then turbo and leave overnight.
1902 Turbo on CVJ = 3.4 ubar.
1903 Zach leaves lab.