2021/03/22 (Monday)

Previous Day

  • Lab in-person: Ruiren
  • Virtual partner: Matt
Status Run 74 in progress. Warming the fridge a bit to pump out helium from IVC, see if that helps the short in DIB2qI that's causing noise.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/HV_diverter_board → DCRC50(3)/55(1). Kepco 0-40V supply → HV BNC. No lead/poly shield.
0830 MB = 33.3%
0845 Tower = 2.26 K, flat overnight. MC heater = 20 mW. Condenser = 68 mbar. 1K Pot = 1.54 K / 6.5 mbar, flat overnight.
We will need to close the 1K Pot speedivalve to get it warmer. But maybe we can not leave it closed too long, so that it won't take so long to pump it out when we restart it.
0850 Turn off 3He pump
1000 Tower temperature starts increasing at this time, breaking out of its plateau. 1K Pot though is still steady. (Maybe turning off the 3He pump weakened the thermal connection through the mixture gas by decreasing the pressure in the fridge.)
1210 Tower = 3.57 K, oscillating in a sorta random way between 3.4 and 3.7 K
1430 Ruiren enters lab
1430 Tower/MC/Coldplate = 3.74K/4.76K/5.79K, Still = 2.98 K / off scale / 0 mW, G1/G2 = 57/51 mbar, 1K Pot = 1.54 K / 7±2 mbar.
1435 Measure short from qI Feedback to chassis: it's still 25 nS.
1440 While I (Matt) watched the short reading on the multimeter, Ruiren shoulder-bumped the cryostat from all four sides. When he bumped it from the East and West sides, I could see the reading drop briefly, to a few nS, then back. So, probably we need to warm up and open the fridge to fix this.
We'll do one more transfer and cool back down to base. Then we can try moving the source, to see how much heating we get. Try to verify that we can move
1455 MB = 23.8%
1530 Begin LHe transfer
1540 Turn off MC heater. Tower temperature falls quickly.
1550 Tower = 2.05 K. 1K Pot = 1.52 K / 7.0 mbar.
1550 Open 1K Pot needle valve → 10°. 1K Pot → 1.52 K / 7.0 mbar.
1555 Open 1K Pot needle valve → 20°. 1K Pot → 1.58 K / 15.0 mbar.
Open 1K Pot needle valve → 30°. 1K Pot → 1.61 K / 22 mbar.
1600 End LHe transfer
1602 MB = 93.9%
1610 Open 1K Pot needle valve → 40°. 1K Pot → 1.64 K / 28 mbar.
1613 MB = 91.2%
1615 Tower = 1.69 K, G1 (fridge) = 41 mbar. (3He pump is off.)
1645 Tower = 1.69 K, G1 (fridge) = 41 mbar. (3He pump is off.) 1K Pot = 1.64 K / 32 mbar.
1650 Turn on 3He pump
1705 Tower = 1.69 K, G1 (fridge) = 48 mbar. 1K Pot = 1.64 K / 30 mbar.
1710 Start circulation through Trap 2. V6 → 11.6%. We'll try this, although last time nothing happened until 12.8%.
1715 Adjust V6 upward in 0.4% steps once a minute. At V6 = 12.8% pressure starts to increase. Condenser peaks at 82 mbar. Tower → 1.67 K
1730 V6 → 13.0%. Condenser peaks at 136 mbar. Tower wavers a bit, → 1.67 K
1755 V6 → 13.2%. Condenser peaks at 270 mbar. Tower → 1.60 K
1830 V6 → 13.4%. Condenser peaks at 450 mbar. Tower → 1.52 K. Still → 150 mbar.
Tower temperature keeps bouncing back up a little after each condenser peak. I know realize we left DCRC3 powered and connected. I should start MIDAS and turn off the 15V power.
1905 Start DAQ programs, then turn off 15V power on DCRC3.
1905 V6 → 13.6%. Condenser peaks at 564 mbar. Tower → 1.43 K. Still → 22 mbar.
1930 V6 → 13.8%. Condenser peaks at 630 mbar (pretty high). Tower → 1.36 K. Still → 10 mbar.
2025 V6 → 14.0%. Condenser peaks at 657 mbar. Tower → 1.32 K. Still → 5 mbar.
2045 Condenser = 558 mbar. Tower = 1.22 K. Still = 3.4 mbar.
2045 V6 → 14.3%. Condenser still falling, → 527 mbar. Tower → 1.14 K. Still → 2.0 mbar.
2055 V6 → 15.0%. Condenser peaks at 526 mbar. Tower → 1.05 K. Still → 0.95 mbar.
2115 V6 → 16.0%. Condenser peaks at 548 mbar. Tower → 0.88 K. Still → 0.50 mbar.
2150 V6 → 18.0%. Condenser peaks at 557 mbar. Tower → 0.83 K. Still → 0.34 mbar.
2210 V6 → 24.0%. Condenser peaks at 565 mbar. Tower → 0.74 K. Still → 0.24 mbar.
2230 V6 → 100.%. Condenser peaks at 567 mbar. Tower → 0.68 K. Still → 0.19 mbar.