start › cdms › k100 › run_diary › 20210319

2021/03/19 (Friday)

Previous Day

  • Lab in-person: Derek
  • Virtual partner: Matt
Status Run 74 in progress.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/HV_diverter_board → DCRC50(3)/55(1). Kepco 0-40V supply → HV BNC. No lead/poly shield.
1414 MB = 61.2%
1500 Tower = 89.5 mK
Terrible noise in DCRC1qI (q4), other channels look fine
1500 Noise in DCRC3qI/qO/DCRC1qI/qO = 1.7/1.4/320/2.3 e-6 @ 10 kHz
1508 Derek enters lab.
1509 Re-tape the transfer stick cap with duct tape.
1510 Try a few things - put 5 V on DCRC1qI only, see noise in DCRC1qO, no change in DCRC1qI. Try turning off 15V power - this seems to sort of turn off both channels. Turn it back on, no improvement in noise.
1513 Tower/MC/Coldplate = 90.7/35.7/180 mK, Still = 1.43 K / 0.130 mbar / 0 mW, G1/G2 = 380/333 mbar, 1K Pot = 1.69 K / 40±2 mbar.
1520 MB = 59.5%
1522 Notice small heating spike around 1510 seems to correspond to when I was on top of the ladder taping up the transfer stick. Wonder if there's a connection…
1526 Fill LN2 cold trap bath.
1527 Tower temperature was falling and now flattens around 90 mK, may be a small bump there as well.
1530 Tower temperature is continuing to fall, was probably nothing more the noise in the thermometer.
1540 Try powercycling DCRC1. No change in noise.
Previously all attempts to fix this sort of noise failed. It always would just go away on its own.
2030 DCRC1qI noise still bad. 420 e-6 @ 10 kHz
0615 DCRC1qI noise still bad. 340 e-6 @ 10 kHz