2021/03/18 (Thursday)

Previous Day

  • Lab in-person: Derek
  • Virtual partner: Matt
Status Run 74 in progress.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/HV_diverter_board → DCRC50(3)/55(1). Kepco 0-40V supply → HV BNC. No lead/poly shield.
0732 MB = 53.5%
0825 Tower = 91.8 mK
0830 Noise in DCRC3qI/qO/DCRC1qI/qO = 1.9/1.3/4.6/1.5 e-6 @ 10 kHz
1432 MB = 42.5%
1448 Derek enters lab.
1450 Tower/MC/Coldplate = 90.4/35.4/186 mK, Still = 1.44 K / 0.130 mbar / 0 mW, G1/G2 = 387/343 mbar, 1K Pot = 1.68 K / 40±2 mbar.
1455 Start 31min series of +10V/-10V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold off/off/80/off on DCRC3qI/qO/DCRC1qI/qO, no in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Start data with Start button on main page
Triggers look fine. Noise doesn't look too bad. No temperature increase like the one we saw yesterday.
Actually there was a temperature increase. It began to increase at 15:00, so 5 minutes after the bias was applied. (Could it actually have been caused by the LN2 transfer??) Tower temperature didn't rise very much, 90.5→92.0 mK over 9 minutes. It peaked there then fell a bit over 3 minutes, then beginning at 15:12 (17 minutes into the series) it rose again, rising to 96.0 mK and leveling off by the end of the series.
1500 Fill LN2 cold trap bath.
1501 MB = 41.8%
1542 Begin LHe transfer @ MB = 41.0%
1608 End LHe transfer
1614 MB = 90.3%