2021/03/02 (Tuesday)

Previous Day

  • Lab in-person: Derek
  • Virtual partner: Matt
Status Run 74 in progress.
Setup: DIB1/2+acc → SL 41/43 → 3.2 MHz filters → cables 2/3 → nothing/nothing → DCRC50(3)/55(1). No lead/poly shield.
1100 Tower = 92.4 mK
1135 MB = 65.2%
Yesterday we disconnected the accessory boards from DCRC1 and plugged in the cable directly. We tried some data but it failed to process - it rejected all randoms from DCRC1, not sure why. But it may be related to very high noise in the channels. This is I think a new problem.
1150 Measure noise with Pulse Viewer in the four channels, looking at 10kHz: DCRC3qI/qO/DCRC1qI/qO = 50/4/150/10 e-6. These are, in order, channels 3/1/4/2. So there is a clear pattern: noise is very bad in the center channel, q4, and gets better as you go out toward the outer rim.
Try zapping SQUIDs to see if that affects the noise. No improvement. Probably the best bet is to powercycle DCRC1.
1200 Apply +5V bias. Noise does not get worse - perhaps slightly better in the two worst channels, q4 and q3. Return bias to 0V.
I didn't do the same measurements before we powercycled DCRC1, but looking at noise traces in BatViewer suggests the q4 noise is 10X what it was before, and q3 is about 80% bigger, and the other two aren't that much worse. So it's suggestive that we have a big noise source in q4 (DCRC1qI) and it's bleeding over into q3 a bit.
1645 Tower/MC/Coldplate = 92.7/31.1/185 mK, Still = 1.42 K / 0.130 mbar / 0 mW, G1/G2= 211/203 mbar, 1K Pot = 1.67 K / 36±2 mbar.
1645 MB = 57.3%
1646 Fill LN2 cold trap bath.
1700 While Derek was in the lab, we were planning to troubleshoot the DCRC noise. But as soon as we looked at it, the noise was much lower than before - it just fixed itself! Looking at 10kHz on the Pulse Viewer again, DCRC3qI/qO/DCRC1qI/qO = 1.3/1.0/3.2/1.1 e-6. This is I think very close to what we saw before. Note that again noise is highest in q4 and gets smaller as you go out - pretty sure we were seeing this before.
OK let's try some data.
1704 Start 31min series of +7V data (bias applied after 1 min), with 1000 beginning randoms, trigger threshold 80/80/80/80 on DCRC3qI/qO/DCRC1qI/qO, no in-run randoms, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
Use sequence r74_takeandflash_dcrcbias.msl
Triggering looks fine.
Turns out this data looks nearly identical to the data with the diverter boards present. No 60keV line visible, muon blob in the same place.