| Status | End of Run 73 warmup. 1K Pot pumped out and isolated. Fridge and 4He and LN2 Trap 1 isolated, not pumped out. N2 gas in IVC. Pumping out cryostat vacuum jacket. |
| 1030 | MC(Cernox) = 285.4 K. tau_heating is now around 102 hours, increased from around 35 hours now that the CVJ is pumped out, good. |
| | Derek's Body Temperature 98.0 ºF |
| 1540 | Derek enters lab |
| 1550 | Turbo on CVJ = 30 nbar. With Turbo valved out for 1 minute, CVJ = 220 nbar |
| 1555 | Stop pumping out CVJ. Attach turbo pump manifold hose to IVC port |
| 1600 | MC(Cernox) = 285.5 K |
| 1605 | Check pressure in IVC: 0.4 bar on the mechanical gauge |
| 1620 | Start removing MB screws to lower MB |
| | Removing the pit covers was not easy as a 1-person job. Used a long 2×4 board, slid it between both handles, so that the board extended to the concrete floor on both sides preventing the possibility of the lid falling into the pit. Then worked to get the lids out. The first one could be rested on the second, but taking the second one off was more awkward. |
| 1710 | Begin lowering MB |
| 1725 | Finish lowering MB into pit |
| 1730 | MC(Cernox) = 286.0 K. At this point temperature starts rising much more quickly, <0.1 K/hr before lowering MB, 2.3 K/hr after. |
| 1735 | Derek exits lab |