start › cdms › k100 › run_diary › 20201111

2020/11/11 (Wednesday)

Previous Day

  • Lab in-person: Ruiren
  • Virtual partner: Matt
Status End of Run 73 warmup. 1K Pot pumped out and isolated. Fridge and 4He and LN2 Trap 1 isolated, not pumped out. N2 gas in IVC and cryostat vacuum jacket.
0800 MC(Cernox) = 281.9 K. It reached dew point at 4:30 am (assuming dew point is 279 K = 42.5 F)
Ruiren temperature = 36.6 C
1110 Ruiren enters lab
1115 Check pressure in cryostat vacuum jacket: 190 μbar. Start pumping out CVJ with Scroll pump then switch to Turbo pump. Turbo → 12 μbar and falling.
1120 Ruiren leaves lab