start › cdms › k100 › run_diary › 20201017

2020/10/17 (Saturday)

Previous Day

  • Lab in-person: Michael
  • Virtual partner: Matt
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs on. Lead/poly shield erected (4 lead walls, 2 poly walls).
0800 MB = 30.7%
0915 Tower = 58.7 mK
0915 Start a 60 min series at HV = 0V,
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pD threshold = 8
Use sequence SimpleDataLoopNoFlash.msl \\Use QETbias_update.c, function sequence_update_QETbiases(62,30)
Tower stayed between 58.4 and 58.7 mK throughout series. But 48 minutes into series, Michael started looking at the other temperatures - this messed up the QET bias updater, so after that point the data will be bad because the QET biases will be too low or too high.
1000 Michael arrives in lab. Body temp 98.7
1005 Tower/MC/Coldplate = 58.6/30.3/195 mK, Still = 1.5 K / 0.120 mbar / 0 mW, G1/G2= 238/229 mbar, 1K Pot = 1.62 K / 22±2 mbar.
1035 Begin LHe Transfer at MB = 26.8%
1040 Fill LN2 cold trap bath
1114 End LHe transfer @ MB = 89.9%
1130 MB = 90.0%
1135 Michael leaves lab
1150 Phonon channels look a little screwy
1150 Zap DCRC1&4 SQUIDs
1540 Tower wandering between 62 and 64 mK. Relock several phonon channels. After that looks like temperature is coming down.
1610 Tower = 59.9 mK and cooling
1636 Start a 60 min series at HV = +300V, drifting 0.0262 V/s, prebias 15% for 2min,
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pD threshold = 8
Use RootSequencer.c, function drift_sequence(300,0.0262)
Use QETbias_update.c, function sequence_update_QETbiases(70,30)
1636 Vtest = 2.5 mV shortly after drift begins. Drift starts at 313.4 V according to RootSequencer. 0.32% HV loss.
Tower wandered between 59.9 and 60.5 mK during series.
1737 Vtest = 2.9 V at end of drift. Final HV = +407.7 V according to RootSequencer. 0.28% HV loss.
1937 Start a 60 min series at HV = +100V, drifting 0.0274 V/s, prebias 15% for 2min,
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pD threshold = 8
Use RootSequencer.c, function drift_sequence(100,0.0274)
Use QETbias_update.c, function sequence_update_QETbiases(70,30)
1937 Vtest = 0.8 mV shortly after drift begins. Drift starts at 107.8 V. 0.30% HV loss.
Tower wandered between 59.8 and 60.5 mK during series.
2038 Vtest = 1.3 V at end of drift. Final HV = +205.4 V. 0.25% HV loss.
2203 Start a 60 min series at HV = +10V, drifting 0.0220 V/s, no prebias,
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = set automatically to track temperature, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR & 5 Hz in-run randoms, trigger on DCRC1pD threshold = 8
Use RootSequencer.c, function drift_sequence(10,0.022,3600,0,0)
Use QETbias_update.c, function sequence_update_QETbiases(62,30)
2203 Vtest = 0.2 mV shortly after drift begins. Drift starts at 11.0 V. 0.73% HV loss.
Tower wandered between 58.8 and 59.4 mK during series.
2304 Vtest = 0.57 V at end of drift. Final HV = +90.2 V. 0.25% HV loss.