start › cdms › k100 › run_diary › 20200924

2020/09/24 (Thursday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Zach, Ruiren
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → pi-filters 1/2/3 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)-master; DCRC33(2) → HV control. DCRCs on. Lead/poly shield erected (4 lead walls, 2 poly walls).
Status Doing a partial warming to get helium out of the IVC. Circulation and 1K Pot stopped. Helium exchange gas in IVC (no gas in cryostat vacuum jacket). MC and Still heaters on and LEDs on to help warm up tower. Bath is slowly boiling away. 1K Pot speedivalve and needle valve open
0016 MB = 11%, Recovery = 383907.6 / 16.3 cuft/hr
0116 MB = 9.5%, Recovery = 383923.9 / 16.3 cuft/hr
0216 MB = 8.1%, Recovery = 383940.2 / 16.3 cuft/hr
0316 MB = 6.7%, Recovery = 383956.4 / 16.2 cuft/hr
0416 MB = 5.3%, Recovery = 383972.7 / 16.4 cuft/hr
0516 MB = 3.9%, Recovery = 383989 / 16.3 cuft/hr
0616 MB = 2.5%, Recovery = 384005.3 / 16.3 cuft/hr
0716 MB = 1.2%, Recovery = 384021.6 / 16.3 cuft/hr
0816 MB = 0.1%, Recovery = 384037.7 / 16 cuft/hr
0916 MB = -0.7%, Recovery = 384054.3 / 16.7 cuft/hr
1016 MB = -1.0%(minimum reading), Recovery = 384070.8 / 16.4 cuft/hr
1116 Recovery = 384088.3 / 17.5 cuft/hr
1216 Recovery = 384105.1 / 16.8 cuft/hr
1035 MC(Cernox)/1K Pot/Sorb = 4.90/5.24/6.6 K, G1(fridge+dump) = 582 mbar. Temperatures were pretty flat through the night but now rising.
1116 Recovery = 384088.3 / 17.5 cuft/hr
1216 Recovery = 384105.6 / 17.3 cuft/hr
1245 MC(Cernox)/1K Pot/Sorb = 5.08/5.77/7.3 K, G1(fridge+dump) = 587 mbar.
1316 Recovery = 384122 / 16.4 cuft/hr
1416 Recovery = 384138.3 / 16.3 cuft/hr
1400 MC(Cernox)/1K Pot/Sorb = 5.15/6.03/7.6 K, G1(fridge+dump) = 589 mbar.
1435 Matt arrives in lab
1440 IVC = 2.5 ubar. Yesterday Zach left it at 1.6 ubar.
1450 Begin pumping out IVC with Turbo Pump. After a minute or so, can see acceleration in the warming of MC and 1K Pot.
1500 Matt Fritts body temperature = 98.1 F
1500 MC(Cernox)/1K Pot/Sorb = 5.41/6.64/8.5 K, all rising. G1(fridge+dump) = 592 mbar.
1516 Recovery = 384153.9 / 15.6 cuft/hr
1530 Fill LN2 cold trap bath
1540 MC(Cernox)/1K Pot/Sorb = 7.18/>7/11.7 K. G1(fridge+dump) = 581 mbar (it fell a little after the LN2 fill).
1545 Matt leaves lab
1615 MC(Cernox)/Sorb = 8.69/13.4 K.
1616 Recovery = 384169 / 15.1 cuft/hr
1710 MC(Cernox)/Sorb = 12.6/16.4 K.
This is about as warm as we got to for the last partial warmup, which was successful. But this time the IVC walls are still cold because there is still LHe in the MB.
1715 Turn off LEDs. This does not seem to slow the warming much. We'll leave them off. Change width/rate from current settings (500/1000 us) to standard settings (100/10000 us).
1720 MC = 13.1 K. Turn off MC&Still Heaters. MC temperature levels out.
1716 Recovery = 384183.8 / 14.9 cuft/hr
1735 MC(Cernox)/Sorb = 12.9/16.2 K. Now that heaters are off these are falling. Interesting that Sorb is affected, somehow MC or Still heater was warming it as well.
1810 MC(Cernox)/Sorb = 12.9/16.0 K. They're mostly flat now.
1816 Recovery = 384198.4 / 14.5 cuft/hr
1916 Recovery = 384212.5 / 14.1 cuft/hr
2016 Recovery = 384226 / 13.5 cuft/hr
2116 Recovery = 384239.1 / 13 cuft/hr
2145 MC(Cernox)/Sorb = 13.2/22.8 K. Now rising.
2216 Recovery = 384251.3 / 12.3 cuft/hr
2316 Recovery = 384262.6 / 11.3 cuft/hr