2020/07/20 (Monday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: everyone at group meeting
0845 Matt Fritts body temperature = 97.2 F
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)-master; DCRC33(2) → HV control. DCRCs on overnight. Lead/poly shield erected (4 lead walls, 2 poly walls).
0750 MB 62.0%
0820 Tower = 68.6 mK, G1/G2= 218/205 mbar, Still = 0.160 mbar / 5 mW, 1K Pot = 1.70 K / 50±3 mbar.
This doesn't look great. Tower temperature not falling, wandered slowly overnight between 66 and 70 mK. Circulation pressures have gone up a lot, though they seem to be leveling out. The relation between G1 and G2 has gone back to normal (G1>G2)
I wonder a little if the mixture is too low - it's gone down slowly over a few years, close to 700 mbar now. One idea is to add a little more 4He. But there is nothing obvious suggesting that it's too low in the still.
0940 Set DCRC50(3) to same settings as DCRC55(1), because we switched to DCRC50(3) yesterday (because DCRC55(1) would not power on)
Note: accessing MIDAS remotely by opening firefox from a putty session on cdms1
Better method: with a Remote Desktop connection to the Fridge computer, you can do everything: access MIDAS from the browser and also use the Pulse Viewer through a Putty connection (with x-forwarding) to CDMS2. The response of the Pulse Viewer is much faster this way.
1000 After some fiddling, get the DAQ programs started on MIDAS (dcrc_driver01, towerfe3_01, triggerfe2).
1005 Start Pulse Viewer, see some noise traces on some channels. Didn't see any pulses.
Detector Side1 channel map
DCRC channel: A B C D
HVbox1 pE pB pD DCRCX(4)
HVbox2 pA pC pF DCRC50(3)
1300 Matt arrives in lab
1300 Tower/MC/Coldplate = 68.8/33.0/202 mK, Still = 1.61 K/ 0.16 mbar / 5 mW, G1/G2= 230/214 mbar, 1K Pot = 1.69 K / 48±3 mbar.
1320 Turn off Still Heater. We'll watch MC thermometer and see what happens.
1350 Turning off Still Heater didn't seem to make things better. MC went down from 33.0→32.2 mK at first then rose to 34.4 mK then settled down to 33.4 mK.
1350 Turn on Still Heater to 5 mW again.
1405 MC spiked at first but now back to where it was an hour ago. Fridge pressures back to what they were too.
1435 Fill LN2 cold trap bath
1440 Matt leaves lab
1620 ZAP all SQUIDs on DCRC 3,4
1700 Try tuning SQUID DCRC3pB. Didn't help clean up its jagged looking pulses
1713 Start a 60 min series at HV = 0,
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (500/360/430)/(360/360/400) uA, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC1pD threshold = 10
Use sequence SimpleDataLoopNoFlash.msl
note: TES settings have not been optimized. This is just a quick dataset so we can make sure to get the processing working. At least Side1pF(DCRC3pD) is looking pretty good.
1725 Tower = 63.4 mK (Tower temperature fell a bit after zapping)
1940 Matt Fritts body temperature = 98.4 F