| 1003 | Matt Fritts body temperature = 97.1 F |
| 1440 | Matt arrives in lab |
| 1450 | Turbo on IVC: 8.6 nbar |
| 1451 | Applied more QCompound. |
| 1530 | Leak checks. @KF-Flange: Test 1 55pbarL/s > 64pbarL/s, Test 2 64pbarL/s > 170pbarL/s, Test 3 91pbarL/s > 180pbarL/s. @Elbow from top down: 100pbarL/s > 330pbarL/s. |
| 1545 | Additional QCompound applied. |
| 1555 | Further leak checks still show similar leak strengths. No reduction in leak. |
| | In summary: strongest leak signal is when squirting down at gap between elbow and fridge flange from above. Goes to ~ 300 pbarL/s. Continued applying Q-compound in gap, filling the volume above the elbow and below the fridge flange. Now starting to fill in volume inside of elbow and outside of fridge flange, by applying Q-compound from above and tamping down. Still more to do, next time. Still wondering whether the fridge flange itself leaks right where the elbow is. I'll eventually cover that in Q-compound too. |
| 1610 | Matt Fritts leaving lab. |
| 2255 | Matt Fritts body temperature = 98.2 F |