| 1042 | Matt Fritts body temperature = 97.6 F |
| 1150 | Matt arrives in lab |
| 1207 | Turbo on IVC: 8.6 nbar (this is lower than it has been, lately it had been bottoming out 12 nbar.) |
| 1212 | With Turbo valved out for 5 minutes, IVC = 0.54 ubar. |
| 1241 | Leak check at KF-Flange to stripline elbow. Leak was much improved; rises from 80pbarL/s to 120 pbarL/s. Leak at top of elbow is higher with half second blast of helium rises to 300pbarL/s. |
| 1245 | Applied more Q-Compound at gap at top of elbow. |
| 1326 | Leak check at top of elbow. Still rises to ~300pbarL/s. |
| 1337 | Matt leaving lab. |
| 1738 | Matt Fritts body temperature = 98.2 F |