| Further leak tests |
| signal reported is rise above base, base~20 pbarL/s |
| 1 m of tube length, through needle | top of knob to North | 15 |
| 1 m of tube length, through needle | base of lower knob to North | 23 |
| 1 m of tube length, through needle | top of knob West, between knob and fridge insert | 16 |
| 1 m of tube length, through needle | top of knob to South | 50 |
| 1 m of tube length, through needle | base of lower knob to South | 135 |
| Push in more Q-compound on South side, more between knobs and fridge insert |
| 1 m of tube length, through needle | base of lower knob to South | 20 |
| blast 1 sec from tube | top of knob around South | 330 |
| blast 1 sec from tube | top of KF flange | 530 |
| that is surprising. Try tightening flange a little more. |
| squirt through needle 0.5 sec | top of KF flange | 38 |
| squirt through needle 0.5 sec | top betw KF and elbow | 120 |
| squirt through needle 0.5 sec | bot of KF flange | 3 |
| squirt through needle 0.5 sec | bot betw KF and elbow | 21 |
| this seems to rule out the KF flange, leak still seems to be at elbow. |