Setup Run 72. DIB 1/3/4 → Stripline 43/41/48 → Cable 3/2/1 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1). DCRCs on overnight. No shield.
Overnight: Tower temperature falling at mK/hr, mK/day
0900 Tower/MC/Coldplate = 53.2/35.7/181 mK, Still = 1.27 K / 0.120 mbar / 0 mW, G1/G2= 413/385 mbar, 1K Pot = 2.12 K / 26 mbar. MB = 66.6%
0905 Fill LN2 cold trap bath
1020 Tower temperature increased after LN2 fill, from 53 mK to 64 mK. Why? I guess maybe because the still got a little colder and we have less cooling power. Let's try heating the still.
1025 Set Still Heater to 5.0 mW. Still and Circulation pressures increase, Tower temperature falls.
1030 Start assembling lead/poly shield
It's been kind of hard to see Am peaks in the data, but in Run 70 and previous it got a lot easier with the shield in place, reducing the background.
1135 Tower = 54.7 mK, Still = 0.190 mbar / 5 mW, G1/G2= 492/462 mbar, 1K Pot = 2.18 K / 24 mbar.
1140 East and West shield walls in place. North and South frames for lead attached. (Skipping N/S poly.)
1430 North and South lead in place, lead in all 4 walls (5 1/8-inch layers)
In Run 70 when the shield was erected (3 walls) we drifted HV at 2.64 V/min = 44 mV/s. But that's only a little lower than what we did this run without the shield. In Run 70 without the shield we drifted at 65 mV/s. If we assume the shield reduces the rate by the same ratio as Run 70 then we should try 31 mV/s now.
1440 Use HVscan.py to generate a sequence HVdrift_250V_31mVps_pre10-2_80min.msl. This should bias detector to 250 V after a 2 minute 10% prebias, then drift by 31 mV/s and run for 80 minutes (ending at about 400V). Then flashes 30 seconds and waits 30 minutes.
1451 Start a series at HV = 250→400V, rate 31mV/s, 80 minutes, with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (na/360/430)/(360/420/440) uA, Front-End Gains = (na/3/3)/(5/5/5)
1000 BOR and 5Hz in-run randoms, trigger on DCRC1pD threshold = 9
Use sequence HVdrift_250V_31mVps_pre10-2_80min.msl
1503 Data starts
1505 Vtest = 3.1 mV, Kepco = 235 V, pulses on BatViewer look OK. (Kepco dial reading is lower than what the Fluke meter reads.)
Tower = 53.4 mK
1520 Vtest = 3.3 mV, Kepco = 250 V, pulses on BatViewer look OK.
1555 Vtest = 4.1 mV, Kepco = 305 V, pulses on BatViewer look OK.
Tower = 53.1 mK
series produced 86 dumps
1745 Fill LN2 cold trap bath