2019/09/23 (Monday)

Previous Day

Status: LN2 transfer Friday and more yesterday. Fridge passed throughput test. Electrical checkout mostly OK.
0900 IVC = 2.2 mbar. Begin pumping out IVC with scroll pump. After a few minutes switch to turbo pump.
0910 Start Pfeiffer Leak Detector using scroll pump as the forepump. On Friday we were still having problems using the forepump in the leak detector.
0910 MC(Cernox)/Sorb = 77.29/115.8 K. G1/G2 = 592/538 mbar, slowly rising.
0930 LD signal on IVC = 40/440 pbarL/s with Turbo pump valved in/out
0945 LD signal on IVC = 27/300 pbarL/s with Turbo pump valved in/out
1010 LD signal on IVC = 16/170 pbarL/s with Turbo pump valved in/out. Leave turbo valved out in preparation to blow out LN2.
1025 LD signal on IVC = 220 pbarL/s
1030 Begin blowing out LN2. Main bath (G3) = 604 mbar above 1atm.
1035 Fill LN2 cold trap bath during blowout
1035 LD signal on IVC = 210 pbarL/s
1045 LD signal on IVC = 196 pbarL/s
1105 LD signal on IVC = 164 pbarL/s
1110 Start pumping mixture out of fridge into dump.
1110 G3 (main bath) → 270 mbar above 1atm and continues to fall. Close V1a temporarily: G3 rises. So looks like we still have helium in the cylinder and that the blowout is almost over.
Throughput test procedure, Cold Trap 2:
  1. All mixture in dump, LN2 trap 2 cold, condenser & still valves open (3He pump on - optional)
  2. Open V9, V13b, V12b+speedivalve, V3. Wait for G1 to be close to G2.
  3. Close V12a+speedivalve and V3, open V1
  4. monitor G1
1115 Perform mixture throughput test at MC = 77 K using above procedure with 3He pump on. Let G1 rise to 520 mbar before Step 3, then it falls to 50 mbar, then rises. Throughput = 37 mbar/min.
1120 G3 (main bath) fell to 150 mbar+1atm, then fell more quickly to 78 mbar+1atm, now it looks like the helium cylinder is empty. But probably blowout is finished.
1120 LD signal on IVC = 117 pbarL/s, looks like no leak. Turn off leak detector.
1125 Start pumping out main bath with scroll pump. Takes 15 minutes to fall to 1 mbar.
1150 With MB down to 0.4 mbar, close off scroll pump. Rises to about 1 mbar. Then open 1K Pot Speedivalve and Needle valve (a few turns). MB rises to 7 mbar then falls to 3.5 mbar. Pump out MB and 1K Pot with scroll pump.
1200 Pressurize MB and 1K Pot with helium to 314 mbar above 1 atm. Close both 1K Pot valves.
1205 Open both sides of fridge to Trap 2, open Trap 2 to dump. (V9, V13b, V12b, V1, V3, Condenser, Still valves open.) 3He pump on.
1210 Let 46 μbar He exchange gas into IVC.
1220 Start recording pictures of recovery meter
1225 MC(Cernox) = 77.3 K, Sorb = 118.1 K, IVC = 46 μbar, G1(fridge) = 604 mbar. G3(MB) = 113 mbar+1atm.
1230 Begin LHe transfer
1230 G3(MB) → 950 mbar+1atm
1240 MC(Cernox) = 76.6 K, Sorb = 115.8 K, IVC = 34 μbar, G1(fridge) = 601 mbar. G3(MB) = 384 mbar+1atm. Motor Coil A = 6.34 kΩ.
MB pressure and recovery rate have settled. Let's start pressurizing the transfer Dewar.
1255 MC(Cernox) = 62.1 K, Sorb = 69.1 K, IVC = 74 μbar, G1(fridge) = 571 mbar. G3(MB) = 350 mbar+1atm. Motor Coil A = 6.32 kΩ.
why is IVC pressure increasing? I had pumped it down a little from a somewhat too high pressure. So maybe it's still settling? This is a bit concerning though.
close valve on pump manifold that leads to the hoses for the scroll pump and leak detector. Maybe there's a leak there.
1300 IVC = 100 μbar. Let's pump it down again, to 63 μbar.
1310 MC(Cernox) = 49.5 K, Sorb = 8.0 K, IVC = 86 μbar, G1(fridge) = 472 mbar. G3(MB) = 391 mbar+1atm. Motor Coil A = 6.27 kΩ.
1315 MC(Cernox) = 44.7 K, Sorb = 7.7 K, IVC = 120 μbar, G1(fridge) = 472 mbar. G3(MB) = 372 mbar+1atm. Motor Coil A = 6.24 kΩ.
Pump IVC down a little more, to around 100 μbar
1340 MC(Cernox) = 22.3 K, Sorb = 6.4 K, IVC = 140 μbar, G1(fridge) = 454 mbar. G3(MB) = 313 mbar+1atm. Motor Coil A = 6.14 kΩ.
Pump IVC down a little more, to around 100 μbar
1345 MB = 63.1%. Had forgotten to look at this.
1346 MC(Cernox) hits 10K. Start pumping out IVC.
Motor Coil A suddenly goes superconducting, right when it's supposed to: → 32 Ω.
1350 MC(Cernox) = 4.34 K, Sorb = 5.2 K, turbo on IVC = 3.4 μbar, G1(fridge) = 440 mbar. G3(MB) = 313 mbar+1atm. Motor Coil A = 31.8 Ω. MB = 73.0%.
1400 End LHe transfer @ MB = 66.5%
1415 MC(Cernox) = 4.23 K, Sorb = 5.2 K, turbo on IVC = 250 nbar, G1(fridge) = 520 mbar. G3(MB) = 126 mbar+1atm.
1515 MC(Cernox) = 4.24 K, Sorb = 5.2 K, turbo on IVC = 250 nbar, G1(fridge) = 533 mbar. G3(MB) = 118 mbar+1atm.
Pump cold LN2 trap 2 to dump.
1545 G1/P1 = 1.1/0.02 mbar. Close V12B+speedi and V6.
Open V12a+speedi, V13A, and V1. Open gate valve to leave mix exposed to fridge and cold trap 1. Turn off 3He pump.
1550 Take out cold trap 2 and fill LN2 bath.
1815 Begin LHe transfer MB=64.1%
1850 End LHe transfer MB=90.3%

Stepper Motor

1454 Looking at DIB1 on breakout box. Motor Coil A/B = 33.3/31.7 Ω, Slide Pot 1-2 = 4.504 kΩ, Slide Pot 1-3 = 16.20 kΩ, Slide Pot 2-3 = 12.35 kΩ
1457 With the standard copper coils the stepper motors each have resistances of 125 ± 0.5 Ω. It would be good to match this and to have the two coils as similar as possible so we are sending the same current down each. Added a couple 200 Ω trim pots on the Dsub50 breakout board for this purpose.
Attach DIB1 to the Dsub50 breakout board and adjust the pots to get total resistances of Motor Coil A/B = 125.0/125.0 Ω.
1515 Testing stepper motor. MC Cernox = 4.23 K. Attach arduino and motor driver board to DB50 breakout board on DIB1. Power Arduino from USB. Power motor board with 7.5V from Instek Lab Power supply. Power supply says the board is currently drawing 10 mA. Slide Pot 1-2 = 4.504 kΩ.
Connect to Arduino via telnet with command telnet cdms-arduino1.
1522 Send command mv +48 to step by 48 steps = 1 rotation = 1/32 inch. MC rises from 4.23 K to 4.26 K, then falls back down within ~2 min. Slide Pot 1-2 rises while motor is moving, then continues to rise as fridge heats, eventually settles at 4.569 kΩ. Looks like it worked and we guessed the polarity correctly.
1528 Slide Pot 1-2 = 4.570 kΩ, Slide Pot 1-3 = 16.19 kΩ, Slide Pot 2-3 = 12.46 kΩ (measured from Dsub50 breakout panel). Odd, it looks like Slide Pot 2-3 increased also, but maybe it's just the different breakout wiring??
1533 Let's try to move back to 0. Send command mv -48. MC again rises to ~4.26 K, then settles back at 4.23 K. Slide Pot 1-2 initially falls to 4.51 kΩ, then rises to 4.582 kΩ. Slide Pot 1-3 = 16.19 kΩ, Slide Pot 2-3 = 12.48 kΩ.
This is strange, the total resistance has remained unchanged as expected, but the two partial resistances have both increased no matter which direction we move.
Could be that the contact resistance of the pot slider is unstable.
1609 Quit Arduino telnet session with the command qu. Power down Motor (Instek) power supply. Unplug Arduino USB power.
Slide Pot 1-2 = 4.582 kΩ
Step number SP 1-2 (kΩ) SP 2-3 (kΩ) SP 1-3 (kΩ)
0 4.504 12.35 16.20
48 4.570 12.46 16.19
0 4.582 12.48 16.19

Based on bench tests, 1 rotation should move the potentiometer through 1.4% of it's total resistance. So we expect SP 1-2 and SP 2-3 to be changing by 233 Ω per rotation.

1720 Let's assume there is a significant contact resistance on the slide pot. If we take the 3 measurements we have and assume a value for the stripline and other wiring resistances, we can work out the fractional position and the contact resistance.

We have:

<m>R_{13} = 2 R_{SL} + R_T </m>
<m>R_{12} = 2 R_{SL} + R_C + f R_T </m>
<m>R_{23} = 2 R_{SL} + R_C + (1-f) R_T </m>

where <m>R_{C}</m> is the contact resistance, <m>R_{SL}</m> is the stripline resistance, <m>f</m> is the fractional position, and <m>R_{T}</m> is the total pot resistance. From these we can show:

<m>f = 1/2 (1- {R_{23}-R_{12}}/{R_{13}-2R_{SL}} )</m>
<m>R_C = {R_{12} + R_{23} - R_{13} - 2 R_{SL}}/2</m>

If we use the motor coil measurements to estimate the stripline and wiring resistance, we get <m>R_{SL}~16.25 Ohm</m>. Using the expressions above, then we can calculate R_C and f. This appears consistent with the slider not moving but the contact resistance changing…

Step number SP 1-2 (kΩ) SP 2-3 (kΩ) SP 1-3 (kΩ) R_C (Ohm) f
0 4.504 12.35 16.2 311 0.257
48 4.57 12.46 16.19 404 0.256
0 4.582 12.48 16.19 420 0.256