start › cdms › k100 › run_diary › 20190616

2019/06/16 (Sunday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
0510 Feed cat breakfast. Birds woke her up.
0513 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
Around 07:30, fridge computer rebooted - I guess? It was a little strange because I didn't have to log in, but LabVIEW and everything else was off. This was with Remote Desktop.
0839 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1115 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1328 Partially fill LN2 cold trap bath. LN2 dewar runs dry.
1330 Tower/MC/Coldplate = 47.6/33.7/199 mK, Still = 1.3 K / 0.095 mbar, G1/G2 = 150/146 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 26.0%
All measurements taken after a recent flash.
1340 Jiajin has not been responding to keyboard/mouse, though sshing to it works. Restart jiajin.
1355 Commence LHe transfer. MB=26.0%
1440 Cease LHe transfer. MB=92.0%
1529 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1838 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
Note: the HV current leakage of the last two data series is noticeably worse than earlier series. In earlier runs the leakage went 2.5→6 mV; in the next-to-last one it was 3→8 mV, and the last one was 5→11 mV. This is a clear effect of the LN2 running out, so the Ebox purge has stopped. This is not a severe effect on the effective HV - less than a 1% difference. But I'll stop taking data for now rather than let it get worse, we'll wait until we get the purge going again.