2019/06/15 (Saturday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Three walls of the lead/poly shield are up.
0525 Feed cat breakfast. Birds woke her up.
0532 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
0915 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1247 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1350 Fill LN2 cold tap bath
1445 Tower/MC/Coldplate = 41.0/30.4/190 mK, Still = 1.2 K / 0.090 mbar, G1/G2 = 149/152 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 59.6%
1543 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
1930 Looks like DCRC1pD was railed for the last entire sequence.
1945 Adjust DCRC1pD driver offset to unrail it
1946 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl
2220 Start a 90 min series at HV = 180→400V, rate 2.418V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR, 5Hz InRun, and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_180V_cal0612_2p418VPM_prebias_10_2_dur_90min_flashDCRC_1_4.msl