2019/06/10 (Monday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC55(1)
DCRC33(2) → 1kV power supply → HV input

Two walls of the lead/poly shield are up.
0845 Tower/MC/Coldplate = 43.1/30.9/178 mK, Still = 1.02 K / 0.090 mbar, G1/G2 = 160/143 mbar, 1K Pot = 1.64 K / 24 mbar. MB = 37.3%
0850 Fill LN2 Cold trap bath
Let's look at QET bias settings. We'll start a series to take the HV up to 200 V with the standard 014drift to 400 V, but we won't write data. That will give us over an hour at bias to look at the QET settings.
1345 Start the sequence HVdrift_200V_2p64VPM_prebias_10_2_dur_75min_flashDCRC_1_4.msl, then turn off Write data, in preparation for QET bias adjustments and IbIs
Channel I_B (uA) Vout (V) log10(oscillations)
DCRC1 PB 410 -0.969
420 -967
430 -0.962
DCRC1 PC 390 -1.0
400 -0.926
410 -0.861
450 -5.16
470 -5.20
475 -0.571
480 -0.556
485 -0.545
DCRC1 PD 440 -0.957
450 -0.894
460 -0.837
480 -2.83
490 -2.50
500
510 Noise is nasty, 80 kHz at -4.1 (oscil)
DCRC4 PA 410 -0.988
420 -0.909
430 -0.847 -5.3
440 -4.8
450 -4.1
460 -4.1
480 -0.595
490 -0.560
500 -0.537
DCRC4 PB 390 -0.928
400 -0.971
410 -0.894 -5.1
DCRC4 PC 430 -1.01
440 -0.931
450 -0.869 -5.1
470 -2.6
480 -3.0
485 0.974 -4.74
490 0.992 -5.25
495 1.01 -5.7
  • The inner ring channels (DCRC1pC/4pA/4pC) have the worst problems with oscillations. The center channel (DCRC1pD) has some problems, and the outer rings have very little.
  • The recent QET biases were set below the oscillation regime. We can try setting them above and see if there's an improvement.
channel old QET bias new QET bias
DCRC4pA 420 490
DCRC4pB 400 400
DCRC4pC 440 490
DCRC1pB 420 420
DCRC1pC 400 480
DCRC1pD 450 450
1535 Commence LHe transfer. MB=28.0%
1620 Cease LHe transfer. MB=93.3%
1650 Fill LN2 cold trap full.
1653 Start a 75 min series at HV = 200→400V, rate +2.64V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_200V_2p64VPM_prebias_10_2_dur_75min_flashDCRC_1_4.msl
1937 Start a 75 min series at HV = 200→400V, rate +2.64V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_200V_2p64VPM_prebias_10_2_dur_75min_flashDCRC_1_4.msl
2200 Start a 75 min series at HV = 200→400V, rate +2.64V/min with 10% prebias for 2 minutes, flash at end with 30 min wait.
CDMS detector: Driver Gains = 1x, DCRC4/1 QET biases = (490/400/490)/(420/480/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR and 1000 EOR randoms, trigger on DCRC1pB/C/D threshold = 6/6/6.
Use sequence HVdrift_200V_2p64VPM_prebias_10_2_dur_75min_flashDCRC_1_4.msl
All data above is bad, I neglected to turn on triggering and left random rate on at 10 Hz.