2019/06/01 (Saturday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)
DCRC33(2) → 1kV power supply → HV input
0000 Tower = 52.5 mK
0020 MB = 87.0%
0220 MB = 84.4%
0320 MB = 83.0%
0420 MB = 81.6%
0520 MB = 80.2%
0620 MB = 78.8%
0720 MB = 77.4%
0820 MB = 75.9%
1020 MB = 73.1%
1120 MB = 71.7%
1220 MB = 70.2%
1320 Tower/MC/Coldplate = 51.6/31.3/177 mK, Still = 1.01 K / 0.095 mbar, G1/G2 = 196/188 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 68.7%
surprisingly low coldplate and still temperatures, probably these thermometers are acting funny.
Tower temperature continues to fall linearly, slowly
1320 Fill LN2 cold trap bath
1520 MB = 65.8%
1620 MB = 64.3%
1720 MB = 63.0%
1820 MB = 61.5%
1855 Start a 2 min SC noise series at HV = 0V
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (420/370/440)/(420/0/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, 10 HZ IR randoms, trigger on DCRC4pA/B/C threshold = 4/5/5
1858 Start a 2 min Normal noise series at HV = 0V
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (420/370/440)/(420/1000/450) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, 10 HZ IR randoms, trigger on DCRC4pA/B/C threshold = 4/5/5
\\Note: DCRC3 pC was still showing pulses with QET bias = 1000 uA, so it's not totally normal.
1920 MB = 60.1%
2020 MB = 58.6%
2120 MB = 57.1%
2220 MB = 55.6%
2320 MB = 54.2%