2019/05/28 (Tuesday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer ?/3/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)
DCRC33(2) → 1kV power supply → HV input
0000 Tower = 65.7 mK
0815 Tower/MC/Coldplate = 64.2/31.9/199 mK, Still = 1.29 K / 0.095 mbar, G1/G2 = 171/164 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 70.2%
Tower temperature continues to fall linearly, slowly
0950 Turbo on IVC = 0.86 nbar. With Turbo valved out, IVC = 4.0 nbar after a few minutes.
1000 Isolate IVC and turbo pump, then unplug and remove Pfeiffer vacuum gauge from turbo pump manifold for cleaning.
1020 Fill LN2 cold trap bath
1030 MB = 67.0%
1200 During reassembly ignition aid of Pfeiffer vacuum gauge broke. Will attempt to solder back together
1540 MB = 59.6%
1620 Fill LN2 cold trap bath full
1840 Tower/MC/Coldplate = 62.4/34.0/131 mK, Still = 1.30 K / 0.095 mbar, G1/G2 = 166/164 mbar, 1K Pot = 1.63 K / 22 mbar. MB = 55.0%

MC and Coldplate temperatures are clearly wrong here