start › cdms › k100 › run_diary › 20190514

2019/05/14 (Tuesday)

Previous Day

Status Warm after hitting fridge blockages, 3 times. Cryostat vacuum jacket and IVC both pumped out.
Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2
0830 Fill LN2 cold trap bath
0830 Mixture has been open to cold LN2 Trap 1 overnight.
Pump mixture out of Trap 1 into dump.
Removed 10.4 mbar. this is still pretty high…
0845 LN2 Trap 2 is warm. Check contents: 13.2-1.4 = 11.8 mbar on G1. Pump out with scroll pump.
0855 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.4%
0900 Increase V6 → 12.9% to restart flow. G1 increases by ~15 mbar as G2 rises.
1000 Stop cleaning through Trap 2 after 65 min. G2 flow pressure was 570-680 mbar. Pump mixture to dump.
Removed 5.1 mbar.
1010 LN2 Trap 1 is mostly warm. Check contents: 12.1-1.7 = 10.4 mbar on G1. Pump out with scroll pump.
1020 Check fridge pressures: LHe trap/Condenser/Still = 1.8/0.3/0.12 mbar. Pump out with scroll pump.
1035 Pump fridge down lower by pumping it to the dump with the 3He pump. Scroll pump through V2 only got Still down to 0.14 mbar - pumping with 3He pump through V6 gets it down to 0.02 mbar.
Change of procedure: we'll try waiting before introducing the mixture into the fridge. Cool to 77K with fridge empty.
1045 With 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump.
Once P2 < 100 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume to < 1 mbar on P2.
1100 Close off scroll pump, then pressurized bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar.
1110 Pressurize main bath and 1K Pot (both sides) to 375 mbar +1atm on G3, then close 1K Pot speedivalve and needle valve.
1110 Turbo on IVC = 10 nbar. With turbo valved out for a few minutes, IVC = 190 nbar.
1115 Let 5.4 mbar He exchange gas into IVC
1115 MC(Cernox) = 292.3 K, Sorb = >249 K (max), IVC = 5.4 mbar, Level resistor = 0.986 kΩ. DIB4 pD SQUID = 3.312 kΩ.
1120 Begin LN2 transfer
1120 LN2 Trap 2 is mostly warm. Check contents: 5.8-0.7 = 5.1 mbar on G1. Pump out with scroll pump.
1130 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.9%
1135 MC(Cernox) = 291.8 K, Sorb = >249 K (max), IVC = 4.6 mbar, Level resistor = 0.993 kΩ. DIB4 pD SQUID = 3.266 kΩ.
1135 Increase V6 → 13.1% to restart flow. G1 falls by ~5 mbar as G2 rises.
1145 Increase V6 → 13.2% to restart flow. G1 falls by ~2 mbar as G2 rises.
1150 MC(Cernox) = 280.3 K, Sorb = >249 K (max), IVC = 4.0 mbar, Level resistor = 1.006 kΩ. DIB4 pD SQUID = 2.909 kΩ.
1205 MC(Cernox) = 262.5 K, Sorb = >249 K (max), IVC = 3.4 mbar, Level resistor = 1.010 kΩ. DIB4 pD SQUID = 2.593 kΩ.
1215 Increase V6 → 13.3% to restart flow. G1 falls by ~4 mbar as G2 rises.
1220 MC(Cernox) = 238.2 K, Sorb = >249 K (max), IVC = 3.0 mbar, Level resistor = 1.027 kΩ. DIB4 pD SQUID > 10 kΩ, unstable.
1230 Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 575-645 mbar. Pump mixture to dump.
Removed 4.2 mbar.
1235 MC(Cernox) = 214.5 K, Sorb = >249 K (max), IVC = 2.5 mbar, Level resistor = 1.050 kΩ. DIB4 pD SQUID = 2.15 kΩ.
1250 Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 13.2%
1250 MC(Cernox) = 190.3 K, Sorb = >249 K (max), IVC = 2.5 mbar, Level resistor = 1.088 kΩ. DIB4 pD SQUID = 3 kΩ, unstable.
1255 Increase V6 → 13.3% to restart flow.
1305 MC(Cernox) = 167.5 K, Sorb = >249 K (max), IVC = 2.2 mbar, Level resistor = 1.137 kΩ. DIB4 pD SQUID = 1.775 kΩ.
1310 Increase V6 → 13.4% to speed up flow. G1 rises by ~1.5 mbar as G2 rises.
1325 MC(Cernox) = 142.3 K, Sorb = 247.7 K, IVC = 1.9 mbar, Level resistor = 1.221 kΩ. DIB4 pD SQUID = 1.9 kΩ, unstable.
1335 MC(Cernox) = 131.7 K, Sorb = 213.3 K, IVC = 1.9 mbar, Level resistor = 1.255 kΩ. DIB4 pD SQUID = 1.618 kΩ.
1350 MC(Cernox) = 121.0 K, Sorb = 179.9 K, IVC = 1.9 mbar, Level resistor = 1.282 kΩ. DIB4 pD SQUID = 1.563 kΩ.
1355 Stop cleaning through Trap 2 after 65 min. G2 flow pressure was 570-600 mbar. Pump mixture to dump.
Removed 2.4 mbar.
1405 MC(Cernox) = 113.4 K, Sorb = 159.1 K, IVC = 1.6 mbar, Level resistor = 1.299 kΩ. DIB4 pD SQUID = 5 kΩ, unstable.
1410 LN2 Trap 1 is mostly warm. Check contents: 5.8-1.6 = 4.2 mbar on G1. Pump out with scroll pump.
1415 Pack up S101 to send to Rupak for re-fabrication.
1420 MC(Cernox) = 106.2 K, Sorb = 143.5 K, IVC = 1.6 mbar, Level resistor = 1.312 kΩ. DIB4 pD SQUID = 1.441 kΩ.
1430 Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 13.2%
1435 MC(Cernox) = 100.5 K, Sorb = 134.8 K, IVC = 1.6 mbar, Level resistor = 1.321 kΩ. DIB4 pD SQUID = 1.412 kΩ.
1440 End LN2 transfer. Put ping-pong ball cover on main bath exhaust.
1440 Increase V6 → 13.3% to restart flow. G1 falls by ~1 mbar as G2 rises.
1510 Increase V6 → 13.4% to restart flow. G1 falls by ~6 mbar as G2 rises.
1535 Stop cleaning through Trap 1 after 65 min. G2 flow pressure was 575-695 mbar. Pump mixture to dump.
Removed 4.6 mbar.
1555 LN2 Trap 2 is mostly warm. Check contents: 3.6-1.2 = 2.4 mbar on G1. Pump out with scroll pump.
1610 MC(Cernox) = 82.8 K, Sorb = 127.4 K, IVC = 1.4 mbar, Level resistor = 1.261 kΩ. DIB4 pD SQUID = 1.505 kΩ.
1705 LN2 Trap 1 is mostly warm. Check contents: 5.0-0.4 = 4.6 mbar on G1. Pump out with scroll pump.
1735 Set up passive cleaning with both sides of dump and pump contents exposed to Trap 1 and Trap 2.
With 3He pump on, open V9, V4, V12A+speedi, V12B+speedi, V13A, V13B.