2019/05/12 (Sunday)
| Status | Warm after hitting fridge blockages, 3 times. Cryostat vacuum jacket and IVC both pumped out. |
|---|---|
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 |
| have been watching warm Cold Trap 1 pressure on G1 since Fri, looking for evidence of a leak. | |
| 1045 | vuk1 seems to have restarted, so there's no plots of the pressures overnight. Start up the fridge vi's. |
| 1050 | Trap 1 (G1) = 1.8 mbar up from 1.5 mbar yesterday. Maybe a little leak or just ambient pressure difference? |
| 1106 | Pump out trap 1 using 4He pump. |
| 1112 | Check trap 2 contents: 1.3-1.0 = 0.3 mbar on G1. Pump out with 4He pump. |
| 1120 | Put both traps into LN2 bath. Fill LN2 bath full. |
| 1135 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.4% |
| 1140 | Increase V6 → 12.7% to restart flow |
| 1150 | Increase V6 → 13.0% to restart flow. G1 decreases by ~3 mbar. |
| 1209 | Increase V6 → 13.1% to get G2>G1. G1 decreases by ~3 mbar. |
| 1235 | Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-640 mbar. Pump mixture to dump. Removed 5.4 mbar. |
| 1258 | Stop pumping out trap 1 and remove trap 1 from LN2 bath. |
| 1300 | Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 13.1% |
| 1308 | Increase V6 → 13.2% to restart flow |
| 1315 | Increase V6 → 13.4% to get G2>G1. G1 increases by 45 mbar. |
| 1400 | Stop cleaning through Trap 2 after 60 min. G2 flow pressure was 630-680 mbar. Pump mixture to dump. Removed 8.5 mbar. |
| 1420 | Stop pumping out trap 2 and remove trap 2 from LN2 bath. |
| 1425 | LN2 Trap 1 is mostly warm. Check contents: 7.3-1.9 = 5.4 mbar on G1. Pump out with 4He pump. |
| 1445 | Return Trap 1 to LN2 bath. |
| 1450 | Set up passive cleaning with both sides of dump and pump contents exposed to Trap 1. With 3He pump on, open V9, V4, V12A+speedi, and V13A. |
| 1452 | Top off LN2 cold trap. |