2019/05/03 (Friday)
| Status | Warming up after a second blockage. Nitrogen in IVC and cryostat vacuum jacket. |
|---|---|
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 |
| 0925 | LN2 Trap 2 is warm. Check contents: 5.0-1.7 = 3.3 mbar on G1. Pump out with 4He pump. |
| 0925 | MC(Cernox) = 282.1 K warming time constant around 14.1 hours, should hit 10 celsius at 10:40. |
| 0945 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.6% |
| 0950 | Increase V6 → 12.8% to restart flow |
| 1000 | Increase V6 → 12.9% to speed up flow |
| 1035 | Increase V6 → 13.0% to speed up flow |
| 1050 | Stop cleaning through Trap 1 after 65 min. G2 flow pressure was 575-700 mbar. Pump mixture to dump. Removed 3.0 mbar. |
| 1150 | Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.8% |
| 1155 | Increase V6 → 12.9% to restart flow |
| 1205 | Increase V6 → 13.0% to restart flow |
| 1225 | Increase V6 → 13.1% to speed up flow |
| 1310 | Stop cleaning through Trap 2 after 80 min. G2 flow pressure was 570-655 mbar. Pump mixture to dump. Removed 4.3 mbar. |
| 1330 | LN2 Trap 1 is warm. Check contents: 4.7-1.7 = 3.0 mbar on G1. Pump out with 4He pump. |
| 1345 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 13.0% |
| 1345 | MC(Cernox) = 285.7 K. Cryostat vacuum jacket = 12 mbar. Start pumping it out with scroll pump. After about 20 minutes, pressure falls below 50 μbar, so switch to turbo pump. |
| 1355 | Increase V6 → 13.1% to restart flow |
| 1410 | Increase V6 → 13.2% to speed up flow |
| 1450 | Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 625-700 mbar. Pump mixture to dump. Removed 5.8 mbar. |
| 1505 | LN2 Trap 2 is warm. Check contents: 5.7-1.4 = 4.3 mbar on G1. Pump out with 4He pump. |
| 1520 | Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 13.2% |
| 1535 | Increase V6 → 13.3% to speed up flow |
| 1550 | Increase V6 → 13.4% to speed up flow |
| 1630 | Stop cleaning through Trap 2 after 70 min. G2 flow pressure was 570-610 mbar. Pump mixture to dump. Removed 4.7 mbar. |
| 1635 | Turbo on cryostat vacuum jacket = 19 nbar. Jacket = 250 nbar with pump valved out a few minutes. Good enough, stop pumping on cryostat vacuum jacket. |
| 1645 | IVC = 190 mbar on digital gauge, 0.4 bar on mechanical gauge. Start pumping it out gently with scroll pump. |
| 1645 | LN2 Trap 1 is warm. Check contents: 7.5-1.7 = 5.8 mbar on G1. Pump out with 4He pump. |
| 1650 | Start pumping IVC with turbo pump |
| 1715 | LN2 Trap 2 is mostly warm. Check contents: 6.3-1.6 = 4.7 mbar on G1. Pump out with 4He pump. |