2019/04/10 (Wednesday)
| Status | Began warmup yesterday. Mixture in fridge and dump. Mixture pumped out of LN2 traps, traps warm. 1K Pot needle valve open. Main bath boiling away. |
|---|---|
| 0945 | G1 (dump/fridge) = 339 mbar. Tower/1K Pot/Sorb = 2.7/>7/9.9 K. MB = 4.4% |
| 1025 | Recovery = 1 cuft/223 seconds. Pretty steady overnight. |
| 1045 | Turbo on IVC = 1.6 nbar. IVC = 16 nbar with pump valved out. |
| 1050 | Let some helium gas into the IVC, less than a tube length. Tower temperature spikes to 3.2 K then falls again, IVC settles at 19 nbar. I guess it condensed. |
| 1055 | Let more helium gas into IVC, total about 1 tube length. Tower climbs to 3.6 K. Fridge pressure rises rapidly. IVC settles at 630 nbar. Sorb and 1KPot temperatures fall. |
| 1105 | Tower/1K Pot/Sorb = 3.8/4.2/5.3 K, G1(dump/fridge) = 580 mbar and rising. IVC = 540 nbar. |
| 1110 | Let more helium gas into IVC, part of a second tube length. Tower temperature climbs faster. IVC settles at 7.4 μbar. |
| 1125 | Tower/1K Pot/Sorb = 4.23/4.4/5.5 K, G1(dump/fridge) = 630 mbar. IVC = 7.4 μbar. MC(Cernox) = 4.24 K. |
| 1130 | MB = 2.2% |
| 1135 | Recovery = 1 cuft/211 seconds |
| 1140 | Isolate IVC with 7.4 μbar. Connect pump manifold to cryostat vacuum jacket port. |
| 1150 | Recovery = 1 cuft/197 seconds Looks like recovery rate began climbing after letting exchange gas into IVC. |
| 1150 | MC(Cernox)/1K Pot/Sorb = 4.27/4.79/6.0 K, G1(dump/fridge) = 637 mbar. |
| 1155 | Isolate dump, turn on 3He pump, open V9. G2(dump) = 508 mbar. |
| 1200 | Start pumping mix to fridge from dump, on both sides (V3 open). V6 = 13.0%. G1(fridge) falling, G2(dump) rising. MC/1K Pot/Sorb temperatures fall but then recover. |
| 1205 | MC(Cernox)/1K Pot/Sorb = 4.27/4.70/5.9 K, G1(fridge)/G2(dump) = 195/657 mbar. Increase V6→13.5%. Fridge pressures change more quickly, temperatures dip again. |
| 1210 | MC(Cernox)/1K Pot/Sorb = 4.26/4.71/5.9 K, P1(fridge)/G2(dump) = 1.8/717.8 mbar. |
| 1210 | Recovery = 1 cuft/223 seconds Slowing down again. Maybe because of outside activity. |
| 1220 | MC(Cernox)/1K Pot/Sorb = 4.28/4.80/6.0 K, P1(fridge)/G2(dump) = 0.18/720.6 mbar. Increase V6→20.0%. |
| 1225 | MB = 1.2% |
| 1230 | MC(Cernox)/1K Pot/Sorb = 4.28/4.81/6.0 K, P1(fridge)/G2(dump) = 0.045/722.5 mbar. Increase V6→100.0%. |
| 1305 | Recovery = 1 cuft/224 seconds Still slow. The boost from letting exchange gas into the IVC was temporary. |
| 1310 | MC(Cernox)/1K Pot/Sorb = 4.29/4.88/6.1 K, P1(fridge)/G2(dump) = 0.018/725.6 mbar. |
| 1320 | MB = -0.1% |
| 1325 | MC(Cernox)/1K Pot/Sorb = 4.30/4.89/6.1 K, P1(fridge)/G2(dump) = 0.017/726.8 mbar. |
| 1355 | MC(Cernox)/1K Pot/Sorb = 4.28/4.87/6.1 K, P1(fridge)/G2(dump) = 0.017/727.9 mbar. |
| 1400 | Close off fridge and 4He trap, by closing Condenser speedivalve, Still gate valve, and V1. No significant drop in pressures so fridge was almost certainly empty. |
| 1410 | MC(Cernox)/1K Pot/Sorb = 4.28/4.88/6.1 K, G2(dump) = 728.2 mbar. |
| 1412 | MB = -0.7% |
| 1420 | Let 2.2 mbar He gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket. After 5 minutes, recovery = 1 cuft/13 seconds, MC/1K Pot/Sorb rising. |
| 1430 | MC(Cernox)/1K Pot/Sorb = 4.44/5.11/6.4 K, G2(dump) = 728.9 mbar. Recovery = 1 cuft/14 seconds. MB = -1.0% (minimum). |
| 1440 | MC(Cernox)/1K Pot/Sorb = 4.58/5.53/7.0 K, G2(dump) = 729.1 mbar. Recovery = 1 cuft/18 seconds. |
| 1450 | MC(Cernox)/1K Pot/Sorb = 4.71/5.83/7.4 K, G2(dump) = 729.1 mbar. Recovery = 1 cuft/24 seconds. |
| Barometric pressure = 30.00“ (source is https://w1.weather.gov/data/obhistory/KMSP.html.) G2 = 729.1 mbar. Corrected dump pressure = 731.8 mbar. A few mbar lower than Run 68 but a few mbar higher than Run 67. | |
| 1500 | MC(Cernox)/1K Pot/Sorb = 4.93/6.22/7.9 K, G2(dump) = 729.0 mbar. Recovery = 1 cuft/26 seconds. |
| 1505 | Dump pressure basically flat. Stop pumping to dump. Close all IGH valves, turn off 3He pump. |
| 1515 | MC(Cernox)/1K Pot/Sorb = 5.40/>7/9.0 K, Recovery = 1 cuft/28 seconds. |
| 1530 | MC(Cernox)/Sorb = 5.94/10.0 K, Recovery = 1 cuft/34 seconds. |
| 1545 | MC(Cernox)/Sorb = 6.81/11.1 K, Recovery = 1 cuft/38 seconds. |