2019/04/01 (Monday)

Previous Day

Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)
DCRC33(2) → 1kV power supply → HV input


Lead poly shield is erected, N/E/W walls complete, no poly or lead on S wall.
0658 Start a background series at HV = 300→400V @ 3.9 V/min (26 min duration) with 1 minute 15% prebias
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (430/410/500)/(500/460/420) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC4pA/B/C threshold = 7/6/6
use sequence looptest_HVdrift_fast_300V_3p9VPM_prebias_15_1_dur_26min.msl
0835 Tower/MC/Coldplate = 39.8/29.8/181 mK, Still = 1.60±0.15 K / 0.150 mbar, G1/G2 = 281/259 mbar, 1K Pot = 1.78 K / 13 mbar. MB = 37.3%
0850 Fill LN2 cold trap bath.
0920 Try closing inlet valve to 4He pump for a few seconds. Spike in 1K Pot pressure and temperature, then it recovers immediately. Still spikes from 0.15 to 0.17 mbar and back. Slight spike in Tower temperature, rises about 0.5 mK then falls back.
1515 Start LHe transfer @ MB = 27.7%
1555 End LHe transfer @ MB = 93.4 %
1706 Fill LN2 cold trap partially. Need to refill LN2 dewar.
1730 MB = 91.5 %