Setup DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)
DCRC33(2) → 1kV power supply → HV input


Lead poly shield is erected, N/E/W walls complete, no poly or lead on S wall.
1344 Tower/MC/Coldplate = 39.8/29.9/188 mK, Still = 1.63±0.15 K / 0.150 mbar, G1/G2 = 276/248 mbar, 1K Pot = 1.79 K / 13 mbar. MB = 30.2%
1354 Start LHe transfer, MB=30.2%
1407 Fill LN2 cold trap full.
1430 Cease LHe transfer, MB 92.5%
1554 Start a background series at HV = 300→400V @ 3.9 V/min (26 min duration) with 1 minute 15% prebias
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (430/410/500)/(500/460/420) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC4pA/B/C threshold = 7/6/6
use sequence looptest_HVdrift_fast_300V_3p9VPM_prebias_15_1_dur_26min.msl
1655 Sequence completed but files aren't being copied to /byseries. Not sure why because LazyLogger seems to be running.
1705 Kill and restart MIDAS. Looks like lazylogger is moving files.
1710 Start a background series at HV = 300→400V @ 3.9 V/min (26 min duration) with 1 minute 15% prebias
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (430/410/500)/(500/460/420) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC4pA/B/C threshold = 7/6/6
use sequence looptest_HVdrift_fast_300V_3p9VPM_prebias_15_1_dur_26min.msl
1926 Start a background series at HV = 300→400V @ 3.9 V/min (26 min duration) with 1 minute 15% prebias
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (430/410/500)/(500/460/420) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC4pA/B/C threshold = 7/6/6
use sequence looptest_HVdrift_fast_300V_3p9VPM_prebias_15_1_dur_26min.msl
2257 Start a background series at HV = 300→400V @ 3.9 V/min (26 min duration) with 1 minute 15% prebias
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (430/410/500)/(500/460/420) uA, Front-End Gains = (2/2/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC4pA/B/C threshold = 7/6/6
use sequence looptest_HVdrift_fast_300V_3p9VPM_prebias_15_1_dur_26min.msl