2018/09/27 (Thursday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 450(gain=5, shaping = 5 us integration, Neg input) → x10/x10 attenuator → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: background -125V data overnight, tower = 30.0 mK
0940 Tower/MC/Coldplate = 30.0/30.6/184 mK, Still = 0.0 mW / 1.3±0.1 K / 0.090 mbar, G1/G2 = 124/124 mbar, 1K Pot = 1.60 K / 18 mbar, Turbo on IVC = 0.86 nbar, MB = 72.2%
0945 Fill LN2 cold trap.
1005 Stop sequence after a flash. Need to fill LN2 Dewar so we should wait till that's done before we bias the detector again.
1145 Move PuBe source into lab. Place barrel at South wall to do NaI calibration.
1305 Remove 22Na source from safe and place on south shield wall for NaI calibration.
1340 Move PuBe source barrel to its standard position behind lead wall outside of the west shield wall
1400 Return 22Na source to safe.
1405 After monkeying around with the NaI setup, determine that the optimal arrangement is the following: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 450(Neg input, gain=5, shaping = 5 us integration, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA This is what we'll use from here on out.
1419 Start 8x2hr PuBe series at HV = -125 V with 15% prebias for 5 min, split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-125×8-15_5PreBias-DCRC2_250V-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10)
NaI detector: 1.1 kV (for real) → NaI B/C (dynodes terminated) → Ortec 450(Neg input, gain=5, shaping = 5 us integration, output=+10V)/ Ortec 575A(Neg input, gain=5, shaping = 3 us) → x10/x10 attenuator → DCRCX(4) PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9
Note the correct NaI bias and the new NaI amplifier arrangement.
1655 Fill LN2 cold trap bath

Next Day