start › cdms › k100 › run_diary › 20180925

2018/09/25 (Tuesday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 450(gain=5, shaping = 5 us integration, Neg input) → x10/x10 attenuator → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: PuBe -125V data overnight, tower = 29.7 mK
1115 Tower/MC/Coldplate = 29.7/30.3/182 mK, Still = 0.0 mW / 1.2±0.1 K / 0.095 mbar, G1/G2 = 136/130 mbar, 1K Pot = 1.60 K / 19 mbar, Turbo on IVC = 0.86 nbar, MB = 70.6%
1120 Fill LN2 cold trap.
1303 Start 20 min NaI test series with NaI amplifiers swapped during cooldown period.
NaI detector: 1.1kV → NaI B/C → Ortec 450(gain=5, shaping = 5 us integration, Neg input) / Ortec 452(gain=5, shaping = 6 us, output=10V, Neg) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC4pA/B threshold = 5/5
1325 Switch NaI amplifiers back to: NaI detector: 1.1kV → NaI B/C → Ortec 450(gain=5, shaping = 5 us integration, Neg input) / Ortec 452(gain=5, shaping = 6 us, output=10V, Neg) → x10/x10 attenuator → DCRC4PB/PA before HV data taking resumes.
not noted at the time, but it seems that we must have removed the PuBe source from the lab at this time
1555 Remove 22Na source from safe and place inside lead/poly shield at bottom of NW corner.
1600 Fill LN2 cold trap full.
1628 Start 8x2hr Na series at HV = -125 V with 15% prebias for 5 min, split into 10 min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-125×8-15_5PreBias-DCRC2_250V-30Flash-30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = (10/10/10)/(10/15/10)
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Neg) / Ortec 450(gain=5, shaping = 5 us integration, Neg input) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/C/D threshold = 7/10/9