UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180825

2018/08/25 (Saturday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: link to /data/cdms1 broke yesterday, so no overnight data
1200 Rename the broken symlink /data/chocula/k100/midasraw to midasraw_temp, then create a real link on chocula called /midasraw with subdirectories /byseries and /livedata. This should allow us to take data and store it on chocula for the time being.
1210 Start 1 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
aborted
1230 Something is apparently wrong with the data taking. Files are being written but they are very small and don't open with BatViewer. Will need to investigate.
1240 Noticed that the DAQ programs (dcrc driver, tower fe, trigger fe) weren't on. Stop sequence and reduce HV from 40 to 0 V.
1242 Start 1 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×1-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1256 Data is working but DCRC3pC is railed low. Try a simple relock
1304 DCRC3pC is back but baseline is higher than I like. Change driver offset from -0.25 → -0.15
1314 DCRC3pC baseline still kinda high. Change driver offset from -0.15 → -0.05
1600 Tower/MC/Coldplate = 30.3/29.1/179 mK, Still = 1.2±0.1 K / 0.070 mbar, G1/G2 = 111/109 mbar, 1K Pot = 1.61 K / 20 mbar, Turbo on IVC = 1.0 nbar, MB = 28.3%
1605 Fill LN2 cold trap bath
1609 Begin LHe transfer, MB = 28.3%
1615 Move NaI attenuators to Ortec amplifier inputs.
1617 Start 30min NaI calibration dataset
NaI detector: 1.1kV → NaI B/C → x10/x10 attenuator → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC4pA/B threshold = 7/5
1621 Not seeing a ton of pulses. Increase trigger thresholds to DCRC4pA/B threshold = 10/10
1655 Move NaI attenuators back to Ortec amplifier outputs.
1700 Cease LHe transfer, MB = 92.2%
1711 Start 7 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×7-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9

Next Day