2018/08/24 (Friday)

Previous Day

Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took SiHV+NaI PuBe data overnight.
0845 Tower/MC/Coldplate = 30.3/29.0/179 mK, Still = 1.3±0.1 K / 0.070 mbar, G1/G2 = 116/119 mbar, 1K Pot = 1.61 K / 20 mbar, Turbo on IVC = 1.0 nbar, MB = 74.5%
0850 Fill LN2 cold trap bath
1246 Shift PuBe source south until it is centered just south of the Cryostat lead shield's south wall. Source is now near south-west corner of fridge frame.
1248 Start 1 hr NaI calibration dataset Looking for PuBe gamma structure
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → x10/x10 attenuator → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC4pA/B threshold = 2/2
1400 The Ortec amplifiers may be saturating at a few MeV. Still evaluating, but let's try moving the attenuators to the input of the amplifiers and take one more spectrum in this mode.
1402 Start 1 hr NaI calibration dataset Looking for PuBe gamma structure
NaI detector: 1.1kV → NaI B/C → x10/x10 attenuator → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC4pA/B threshold = 2/2
1445 Stop run early. Triggering only on noise. Increase triggers to DCRC4pA/B threshold = 7/5. Also realise NaI B amplifier output was still plugged into the scope, not the DCRC! Fix this.
1451 Start 30min NaI calibration dataset Looking for PuBe gamma structure
NaI detector: 1.1kV → NaI B/C → x10/x10 attenuator → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC4pA/B threshold = 7/5
1600 Return NaI attenuators to amplifier outputs.
1630 Network problem: /data/cdms1 is inaccessible. This is the path to the raw data. We can't process data, and we can't look at data with BatViewer.
Tried taking data with MIDAS with logger on. Data run immediately stops itself. Checked Messages - but this causes MIDAS to crash. I think Messages needs /data/cdms1.
1650 Move PuBe source barrel back to its central position behind the west shield wall.
1655 Fill LN2 cold trap bath

Next Day