Setup:
DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)

1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V) / Ortec 575A(gain=10, shaping = 3us) → DCRCX(4)
DCRC33(2) → kepco 1kV power supply → HV input
status: Took SiHV+NaI PuBe data overnight. DCRC1pB railed during 07180813_1901 so all data after that is bad.
0905 Tower/MC/Coldplate = 32.0/29.0/167 mK, Still = 1.0±0.1 K / 0.080 mbar, G1/G2 = 124/121 mbar, 1K Pot = 1.64 K / 30 mbar, Turbo on IVC = 1.0 nbar, MB = 18.4%
0910 Fill LN2 cold trap bath.
0910 Pause sequence for a few minutes because I thought we were between data taking periods, but actually we are in the middle of one. We'll probably want to stop data to do the LHe transfer.
0933 Start LHe transfer @ MB = 17.9%
1053 End LHe transfer @ MB = 93.7%
1120 Unrail DCRC1pB and DCRC3pB.
1122 Start 4 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×4-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9
1728 Fill LN2 cold trap full.
2200 All channels still look good.
2208 Start 4 x 2hr 40 V PuBe datasets each split into 10min subruns, with CDMS and NaI detectors running and DCRC2 as HV controller. Using sequence HVscan-40×4-15_5PreBias-30Flash 30wait-2hrData-10minSubRun.msl
CDMS detector: Driver Gains = 2x, QET biases = (280/260/305)/(140/280/295) uA, Front-End Gains = 10
NaI detector: 1.1kV → NaI B/C → Ortec 452(gain=5, shaping = 6 us, output=10V, Pos) / Ortec 575A(gain=10, shaping = 3us, Pos) → DCRC4PB/PA, Driver Gains = 1/1, PhononWaveFormLength = 1024 (512 samples), PhononPrepulseSamples = 64 (64 samples)
1000 BOR randoms, trigger on DCRC3pB/D threshold = 7/9