2018/06/21 (Thursday)

Previous Day

overnight: Tower temp, 1K pot steady
0908 Tower/MC/Coldplate = 40.6/39.3/181 mK, Still = 1.1 K / 0.055 mbar / 1 mW, G1/G2 = 264/246 mbar, 1K Pot = 1.46 K / 4.5 mbar. MB = 39.4%
0913 Fill LN2 cold trap full.
0920 We're going to try cleaning the 4He cold trap. Maybe it is partially blocked and reducing our flow rate. Following procedure in sec 8.3.4 of K100 Operator's Handbook.
0924 Turn off Still heater. Close V6 to stop circulation.
0931 Close V12A and cold trap 1 speedivalve. Close condenser speedivalve. Open V5 to pump mix in condenser lines to G2 volume. Tower temps tarts to rise.
0934 G1/G2 = 2.4/251 mbar. P1 = 1.4 mbar. Tower = 187 mK
0942 G1 seems to be at it's minimum. G1/G2 = 0.7/253.6 mbar. P1 = 5.0 mbar. Tower = 290 mK and rising.
0943 Close V5. G1 → 1.4±0.1 mbar
0945 Procedure says to start pumping out the 4He cold trap next before warming it. I'm going to deviate from that and let it warm so we can see the pressure in G1 rise and get a sense of how much gunk it trapped before pumping it out. Remove 4He cold trap from main bath. G1 immediately rises to 73 mbar, then starts slowly rising. Maybe a little air got in one of the seals as I was lifting the trap out? Check both seals on 4He cold trap lines, they're pretty tight.
0953 G1/G2 = 81.8/254.0 mbar. P1 = 8.0 mbar. Tower = 374 mK
1000 Cold trap feels close to room temp now. G1/G2 = 83.1/254.3 mbar. P1 = 12.0 mbar. Tower = 398 mK
1002 Close V4A. Switch 4He pump from recovery to exhaust. Open V5a, V2a, V7, V2 to pump out 4He cold trap.
1006 G1/G2 = 2.0/252.6 mbar. P1 = 26 mbar. Tower = 423 mK. That seems as low as it will go.
1008 Close V2, V7, V2A, V5A. Switch 4He pump exhaust bask to recovery. Open V4A. P2 rises off scale, but quickly recovers to 3 mbar.
1009 G1/G2 = 4.1/252.7 mbar. P1 = 30 mbar. Tower = 444 mK. G2 decreased by 1.3 mbar while pumping out the 4He trap. Maybe there was a little leak past V12A and the cold trap speedivalve?
1012 Try to replace 4He cold trap in main bath. I'm struggling to get the 4He cold trap inserted fully into the bath. With about 5“ still protruding, it feels like I am hitting up against one of the fiberglass baffles and can't get the trap through the hole. I guess I'll just leave it as is. Hopefully it still gets cooled enough.
1022 G1/G2 = 0.8/255.0 mbar. P1 = 34 mbar. Tower = 464 mK.
1025 Open V12A and cold trap 1 speedivalve to let mix from the LN2 cold trap into the condenser line. G1/G2 → 242/225 mbar.
1028 I'm a little worried we got some air in the condenser line when I pulled out the 4He cold trap at 0945. Let's try and cycle the G1 volume through the LN2 cold trap a while before restarting circulation. Open V5 to cycle warm mix through cold trap 1.
1030 G1/G2 → 69/235.
1032 1K pot seems to be running again: 1.47 K / 5.0 mbar.
1039 LN2 trap 1 line is frosted a little bit, but well below the o-ring seal.
1040 G1/G2 = 69.0/235.4 mbar. P1 = 38 mbar. Tower = 504 mK.
1048 Close V5. Hopefully 20 min of such cleaning is sufficient for the small amount of mix out here. G1/G2 = 244/225 mbar.
1049 Open Condenser speedivalve. G1/G2 drop to ~180 mbar. Open V6 → 12.8%
1053 G1/G2 = 192/179 mbar. P1 = 34 mbar. Tower = 502 mK.
1055 Put IGH into remote mode. At vuk-01, increase V6 → 13.0% Check on IGH manually, seems o have worked.
1056 On VNC session to vuk-01, increase V6 → 13.2% Check on IGH manually, seems o have worked.
1100 Increase V6 → 13.6%
1110 G1/G2 = 214/200 mbar. P1 = 2.2 mbar. Tower = 497 mK.
1128 G1/G2 = 193/180 mbar. P1 = 1.6 mbar. Tower = 472 mK.
1130 Increase V6 → 14.0%
1145 G1/G2 = 220/207 mbar. P1 = 0.6 mbar. Tower = 476 mK.
1200 G1/G2 = 239/224 mbar. P1 = 0.65 mbar. Tower = 452 mK.
1216 G1/G2 = 267/249 mbar. P1 = 0.65 mbar. Tower = 418 mK.
1234 G1/G2 = 290/270 mbar. P1 = 0.65 mbar. Tower = 395 mK.
1256 G1/G2 = 303/281 mbar. P1 = 0.70 mbar. Tower = 372 mK.
1343 G1/G2 = 309/286 mbar. P1 = 0.55 mbar. Tower = 270 mK.
1356 G1/G2 = 309/287 mbar. P1 = 0.55 mbar. Tower = 286 mK. Condenser pressure has flattened out and temp bounced back up from a min of 262 mK. Increase V6 → 14.2%
1406 Increase V6 → 14.4%
1415 G1/G2 = 329/305 mbar. P1 = 0.40 mbar. Tower = 290 mK.
1423 Start LHe transfer (MB = 31.0%)
1430 Increase V6 → 14.8%
1440 Condenser pressure falling since the transfer started and tower temp is coming down again. G1/G2 = 310/289 mbar. P1 = 0.26 mbar. Tower = 226 mK.
1500 Cease LHe transfer (MB = 91.2%), G1/G2 start to rise.
1505 G1/G2 = 324/301 mbar. P1 = 0.26 mbar. Tower = 164 mK.
1517 G1/G2 = 331/307 mbar. P1= 0.22 mbar. Tower = 132 mK.
1528 Increase V6 → 15.0%
1538 G1/G2 = 299/278 mbar. P1= 0.18 mbar. Tower = 115 mK. Increase V6 → 15.4%
1545 G1/G2 = 286/266 mbar. P1= 0.14 mbar. Tower = 112 mK. Increase V6 → 15.8%
1607 G1/G2 = 249/231 mbar. P1= 0.11 mbar. Tower = 89.7 mK. Increase V6 → 100%
1620 G1/G2 = 235/219 mbar. P1= 0.036 mbar. Tower = 75.2 mK.
1640 Fill LN2 cold trap full.
1650 G1/G2 = 234/219 mbar and flattened out. P1= 0.040 mbar. Tower = 53.8 mK.
1840 G1/G2 = 315/293 mbar and rising. P1= 0.040 mbar. Tower = 50.9 mK. Tower bottomed out at 46.5 mK now rising slightly.
1844 Set Still heat = 1.0 mW
1924 Tower bottomed out at 44.8 mK, but now 79.3 mK and rising. Turn off Still heat. Looks like the cold trap cleaning did not help things.