| | overnight: Tower temp, 1K pot steady |
| 0917 | Tower/MC/Coldplate = 40.6/38.8/182 mK, Still = 1.1 K / 0.055 mbar / 1 mW, G1/G2 = 267/247 mbar, 1K Pot = 1.48 K / 5.0 mbar. MB = 71.5% |
| 0940 | Fill LN2 cold trap full. |
| 0947 | Begin cooling power test. Optimal base temp has been with Still heater at 1 mW, so we'll leave it as such. |
| 0948 | MC heater = 5 uW. Temp rising |
| 1022 | MC = 64 mK and rising slowly |
| 1034 | Temps are kind of flattening out. MC = 65.6 mK, Tower = 68.2 mK. We don't have all day, so let's get on with it. |
| 1038 | MC heater = 10 uW. Temp rising |
| 1108 | Temps starting to flatten out, MC = 104.5 mK, Tower = 105.6 |
| 1110 | MC heater = 15 uW. Temp rising |
| 1145 | Temps starting to flatten out, MC = 124.2 mK, Tower = 125.1 mK |
| 1147 | MC heater = 20 uW. Temp rising |
| 1207 | Temps starting to flatten out, MC = 132.0 mK, Tower = 125.1 mK |
| 1208 | MC heater = 30 uW. Temp rising |
| 1248 | Temps starting to flatten out, MC = 169.0 mK, Tower = 169.3 mK |
| 1251 | MC heater = 40 uW. Temp rising |
| 1216 | Temps starting to flatten out, MC = 189.2 mK, Tower = 190.2 mK |
| 1322 | MC heater = 60 uW. Temp rising |
| 1358 | Temps starting to flatten out, MC = 266.2 mK, Tower = 277.2 mK. Circulation pressures and 1K pot remained the same throughout this test. |
| 1401 | MC heater = 0 uW. Temp falling |