2018/06/06 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
overnight: 100V sequence with flashing. Last flash at 01:45. Channels are all railed this morning.
1010 Tower/MC/Coldplate = 37.5/33.0/201 mK, Still = 1.43 K / 0.085 mbar, G1/G2 = 130/128 mbar, 1K Pot = 1.36 K / 1.6 mbar. MB = 67.9%
1K Pot very low, looks like it has run dry. Luckily this has not caused a fridge failure. Still and circulation pressures are also low, but base is elevated slightly.
1015 Open 1K Pot Needle Valve about 10 degrees further. 1K Pot → 1.46 K / 0.45 mbar. Still → 0.095. Tower rises 0.5 mK then starts falling.
20 minutes after tweaking the needle valve, the pot pressure&temperature went into oscillatory mode (this is the “good mode”, probably means the pot is drawing in LHe and maintaining a level). 1K Pot pressure/temperature are drifting downward however.
1015 Fill LN2 cold trap bath
1130 Tower = 36.8 mK, Still = 0.100 mbar, G1/G2 = 152/151 mbar, 1K Pot = 1.48±0.01 K / 5.0 mbar, drifting downward.
1145 Start sequence of 4 30-minute background data series at HV = 100V, with 1% prebias for 5 min (remote HV not calibrated above 100V), Driver Gains = 2x, QET biases = 290/270/315/140/315/310 uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 10/20/20, 30s flash, 30min cooldown, using sequencer script DataLoop_100V-1_5PreBias-30Flash-30Wait-30minData.msl. Tower = 36.8 mK.
1150 Look at pulses during prebias time by disabling logger and starting a run manually. Adjust DCRC3 driver offset. Stop run and re-enable logger before sequence finished pre-bias phase.
1240 1K Pot has fallen to 1.38 K / 2.0 mbar and is no longer oscillating. Open needle valve additional 10 degrees. Stronger effect, Pot peaks at 1.7K / 65 mbar, then settles…
1305 1K Pot = 1.55 K / 9.0 mbar, oscillating.
1405 1K Pot = 1.48 K / 5.0 mbar, oscillating. It keeps drifting downward. Not stable, unfortunately.
1425 1K Pot = 1.43 K / 3.4 mbar, oscillating. Keeps falling.
1440 1K Pot = 1.40 K / 2.4 mbar, not oscillating. Open needle valve another 10 degrees further.
1445 Check on data. All DCRC3 channels railed at the beginning of the second series (07180606_1306). Thereafter, the dumps are empty since we're triggering on DCRC3. We are in the third series now (07180606_1420). Use the DCRC control GUI to re-lock DCRC3 channels. Data starts triggering again.
1536 During prebias for final series, DCRC3 channels are railed again. Relock and adjust them using the SQUID tuning GUI before data starts.
1555 1K Pot = 1.52 K / 6.5 mbar, oscillating. Falling, but maybe a little more slowly than earlier today.
1710 Fill LN2 cold trap bath
1710 1K Pot = 1.49 K / 5.0 mbar, oscillating. Pretty low, worried that it won't last the night, so open needle valve additional 10 degrees.