start › cdms › k100 › run_diary › 20180604

2018/06/04 (Monday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
overnight: DCRCs powered, 60V bias on detector
0938 Tower/MC/Coldplate = 34.7/30.6/195 mK, Still = 1.40 K / 0.100 mbar, G1/G2 = 165/155 mbar, 1K Pot = 1.52±0.01 K / 6.5±0.5 mbar. MB = 72.8%
0940 Vbias/Vtest/Vdet = 60.51/4.2/58.8 V
0943 Dial HV down to 0V and ground.
0945 Fill LN2 cold trap bath.
0945 Dial HV down to 0 V
0950 Flash detector (DCRCs 1,3) for 30 seconds
1035 Tower = 34.9 mK
1300 Realized why the Keithley Electrometer was still having issues measuring the test resistor voltage. There was a jumper on the back of the device connecting the “lo” side of the voltage to ground. Disconnect this jumper.
1315 Dial prebias HV up to 55 V.
1320 Vbias/Vtest/Vdet = 55.21/5.48/53.02 V.
1330 Dial HV down, Vbias/Vtest/Vdet = 40.04/1.21/39.56 V
1332 Start sequence of 1 60-minute background data series at HV = 40V, Driver Gains = 2x, QET biases = 290/270/315/140/315/310 uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 10/20/20, no flash between, using sequencer script SimpleDataLoopNoFlash.msl. Tower = 34.8 mK.
1440 Vbias/Vtest/Vdet = 40.04/2.30/39.12 V
1445 Dial HV down to 0 V and short leads.
1455 Unplug variac from wall, plug into power strip connected to the power relay card for the Kepco HV supply. Plug power strip into wall with disconnected ground plug. Plug the power relay card into wall. Plug voltage cable into D50 breakout board connected to DCRC4. Power DCRC4. Turn off DAQ programs, adjust ODB for all three DCRCs, turn DAQ programs on again.
1500 Currently Kepco HV is turned all the way down. Run sequence changeHV_0_70.msl to calibrate remote HV control.
1500 Actually this didn't work, need to turn on HV first by setting DCRC4q[1] to 2 V. Then set DCRC4q[0] to 10.652, which is the “HV=0” value according to the sequence script. Try again. After sequence finishes, dial manually to 70.00 V.
1505 Test HV control by running changeHV_70_0.msl. Ends at -0.18 V. Try turning off HV - now reads 0.86 V.
1550 DCRC3 channels railed. Recover by relocking and adjusting driver offsets.
1648 Fill LN2 cold trap full.
1656 Start HV scan of 60-minute background data series at HV = 0,10,20,30,40 V, with 15% prebias for 5 min, Driver Gains = 2x, QET biases = 290/270/315/140/315/310 uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 10/20/20, flash for 30s, cool for 30 min, using sequencer script HVscan-0_10_20_30_40V-15_5PreBias-30Flash-30wait-1hrData.msl