| | Status: Replaced 1K pot shaft orings. Cleaning mixture. |
| 0854 | IVC pressure = 5.4 mbar |
| 1340 | Pump out fridge lines with scroll pump. |
| 1344 | Let mixture from 3He pump through cold LN2 Trap 1 into fridge lines, both sides. (Most mixture still in dump.) |
| 1350 | Pump out main bath with scroll pump. |
| 1353 | Once P2 < 10 mbar, open 1K pot speedivalve. Open 1K pot needle valve 5 full turns. |
| 1356 | P2 = 0.4 mbar, close V1A and 1K pot speedivalve. |
| 1401 | Open V5A, V2A, vent speedivalve to pump out clear He hose. Close off scroll pump. |
| 1405 | Pressurize main bath and 1K pot with helium to 316 mbar above 1 atm on G3. |
| 1406 | Close 1K pot needle valve. |
| 1413 | Begin LN2 transfer |
| 1420 | MC(Cernox) = 291.8 K, IVC = 5.4 mbar, G1 (fridge) = 211 mbar, New level resistor = 2.163 kΩ |
| 1430 | MC(Cernox) = 290.2 K, IVC = 4.6 mbar, G1 (fridge) = 208 mbar, New level resistor = 2.269 kΩ |
| 1450 | MC(Cernox) = 273.9 K, IVC = 3.4 mbar, G1 (fridge) = 205 mbar, New level resistor = 2.884 kΩ |
| 1505 | MC(Cernox) = 249.8 K, IVC = 3.0 mbar, G1 (fridge) = 203 mbar, New level resistor = 2.996 kΩ |
| 1520 | MC(Cernox) = 230.2 K, IVC = 2.5 mbar, G1 (fridge) = 202 mbar, New level resistor = 3.005 kΩ |
| 1540 | MC(Cernox) = 209.7 K, IVC = 2.5 mbar, G1 (fridge) = 202 mbar, New level resistor = 3.005 kΩ |
| 1600 | MC(Cernox) = 199.4 K, IVC = 2.5 mbar, G1 (fridge) = 200 mbar, New level resistor = 3.005 kΩ |
| 1625 | MC(Cernox)/Sorb = 176.0/183.9 K, IVC = 2.2 mbar, G1 (fridge) = 199 mbar, New level resistor = 3.004 kΩ |
| 1627 | Open LN2 dewar valve open all the way. It was already quite open (~3 turns) though. |
| 1632 | Cease LN2 transfer. Started to hear LN2 splash out of the open main bath port. |
| 1634 | MC(Cernox)/Sorb = 170.2/175.1 K, IVC = 2.2 mbar, G1 (fridge) = 197 mbar, New level resistor = 2.999 kΩ |
| 1800 | Fill LN2 cold trap full |