2018/05/17 (Thursday)
| Status: 1K Pot is not working. Warming up to examine it. | |
| 0540 | MC(Cernox) = 116.9 K |
| 0710 | MC(Cernox) = 125.1 K |
| 0740 | MC(Cernox) = 127.8 K |
| slow warming curve. Maybe the still and MC heaters help, didn't use them this time. | |
| 0905 | MC(Cernox)/1K Pot/Sorb = 135.1/>7/>250 K. Pot and Sorb rising steadily. |
| 0905 | Turn on MC and Still heaters to 20 mW each |
| 0905 | Cryostat vacuum jacket = 100 μbar |
| 0910 | Let two hose-lengths He gas into cryostat vacuum jacket. Vacuum jacket → 1.2 mbar |
| 0910 | Set up heat gun to blow up at bottom of main bath |
| 0915 | Make sure 1K Pot needle valve is fully closed, then check pressure in 1K Pot: 50 mbar on P2, 1.5/30 bar on mechanical gauge. Pump out with scroll pump. |
| 0920 | Stop pumping on 1K Pot, leave speedivalve open to watch pressure on P2: |
| minutes | mbar (P2) |
|---|---|
| 0 | 0.08 |
| 0.5 | 0.45 |
| 1.0 | 1.3 |
| 1.5 | 2.4 |
| 2.0 | 3.4 |
| 3.0 | 4.5 |
| 4.0 | 5.5 |
| 5.0 | 7.0 |
| 0925 | Pump out 1K Pot again with scroll pump, down to 0.080 mbar. Open needle valve a few degrees, → 0.090 mbar. Open needle valve a few more degrees, → 0.110 mbar. Open needle valve to about 90 degrees, → 0.260 mbar Open needle valve to about 180 degrees, → 4.5 mbar |
|---|---|
| 0930 | With needle valve open 180 degrees, stop pumping on 1K Pot, leave speedivalve open to watch pressure on P2: rises off-scale in 20 seconds. After a few minutes reaches 13/30 bar on mechanical gauge, continues to rise. |
| 0950 | Pump out 1K Pot again with scroll pump, down to 0.080 mbar. Open needle valve 10 degrees, → 0.110 mbar. |
| 0950 | With needle valve open 10 degrees, stop pumping on 1K Pot, leave speedivalve open to watch pressure on P2: |
| minutes | mbar (P2) |
|---|---|
| 0 | 0.11 |
| 0.5 | 0.9 |
| 1.0 | 2.8 |
| 1.5 | 4.5 |
| 2.0 | 5.5 |
| 3.0 | 9.5 |
| 0958 | MC(Cernox) = 141.9 K |
|---|---|
| 1200 | MC(Cernox) = 168.3 K |
| 1355 | MC(Cernox) = 190.6 K |
| 1625 | MC(Cernox) = 215.3 K |
| 1710 | MC(Cernox) = 221.4 K |
| 1910 | MC(Cernox) = 236.1 K |