UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180315

2018/03/15 (Thursday)

Previous Day

Status: LN2 transfer yesterday. Several cryocleanings done, mixture exposed to cold LN2 Trap 1 overnight (no mixture in fridge). 1K Pot evacuated and isolated.
0915 MC(Cernox) = 76.9 K, Sorb = 118.7 K, IVC = 3.4 mbar
increase in IVC overnight is worrying
0920 Pump mixture out of Trap 1 to dump
0920 Pump out IVC with scroll pump, then turbo pump
0940 Let about 440 mbar mixture into cold LN2 Trap 2 through v13b, then close v13b
0945 Closer Condenser valve, open v1. Then open 12b+speedivalve, and close it again. This lets the mixture in Trap 2 into the 4He cold trap. Then open condenser valve. This lets the mixture in the 4He trap into the condenser. G1 = 398 mbar before opening Condenser, 363 mbar after opening Condenser.
This in itself is a mini- throughput test. Since the still side is vacuum we should see G1 fall as this mixture moves through the still impedance.
0950 Open Still valve. This should allow us to see the pressure on the still side rise on P1.
1000 Throughput doesn't look good. Condenser is falling very slowly, 0.02 mbar/min. This may simply be the mixture cooling. P1 is very low, 0.022 mbar, rising extremely slowly, 0.2 μbar/min.
1005 Open v3 to let mixture into Still, so condenser side and still side equilibrate. G1 (condenser) falls from 363 mbar to 17 mbar.
1010 Let more mixture from dump through Trap 2 into fridge. G1 → 256 mbar.
1010 Mini-throughput check: close off Trap 2 and V3. Only V1, condenser speedivalve, still gate valve, and V9 open. Turn on 3He pump, then open V6 to 12.8%. This pumps mix out of the still side and we can see if the pressure goes down on the condenser side. G1 (condenser) started out at 256 mbar, did not fall. Throughput < 0.04 mbar/min.
1025 Pump mixture from fridge and Trap 2 to dump
Throughput test procedure, Cold Trap 2:
  1. All mixture in dump, LN2 traps cold, condenser & still valves open (3He pump on - optional)
  2. Open V9, V13b, V12b+speedivalve, V3. Wait for G1 to be close to G2.
  3. Close V12b+speedivalve and V3, open V1
  4. monitor G1
1030 Perform mixture throughput test at MC = 77 K using above procedure. Let G1 rise to 570 mbar before Step 3, then it falls to 32 mbar, rise is very slow. Throughput 0.08 mbar/min.
We have another blockage.
1045 Pump mixture out of fridge and Trap 2 to dump
1240 Leak Detector on IVC. Close off turbo pump. LD = 146 pbarL/s
1240 Put level resistor in main bath: 1.237 kΩ. MC(Cernox) = 76.9 K, Sorb = 119.6 K
1250 LD = 157 pbarL/s
1250 Begin blowing out LN2 into cold trap bath Dewar. G3 (main bath) = 498 mbar. After 5 minutes switch to 25L bomb Dewar.
1255 warm LN2 Trap 1 = 10.0 mbar. This was the one the mixture was exposed to overnight.
1300 LD = 155 pbarL/s. Main bath = 491 mbar (above 1 atm).
1310 LD = 150 pbarL/s. Main bath = 484 mbar.
1320 LD = 148 pbarL/s. Main bath = 477 mbar.
1325 LD = 172 pbarL/s. Main bath = 474 mbar.
1330 LD = 266 pbarL/s. Main bath = 479 mbar.
1335 LD = 346 pbarL/s. Main bath = 245 mbar and falling, blowout almost finished.
1337 LD rising quickly… Main bath = 178 mbar
1340 LD = 350,000 pbarL/s. Main bath = 163 mbar.
this leak is actually worse than last time. It really seems to happen right when the pressure falls, which seems like it's not happening at the level of the IVC joint. So it's happening when we start blowing helium into the lab, maybe??
Closing the IVC valves at the top of the fridge causes the leak rate to fall, so it's not a problem with the downstream plumbing
1345 LD = 740,000 pbarL/s. Main bath = 151 mbar.
1345 Stop pressurizing main bath, blowout finished
1350 LD = 712,000 pbarL/s. Main bath = 104 mbar above 1 atm.
1353 Begin pumping out IVC with scroll pump. Pump a little slower than usual to watch the leak rate.
1355 LD = 600,000 pbarL/s. Main bath = 22/30 bar absolute (mechanical gauge).
1400 LD = 390,000 pbarL/s. Main bath = 10/30 bar.
1405 LD = 50,000 pbarL/s. Main bath = 10 mbar (P2).
1410 LD = 6,000 pbarL/s. Main bath = 0.36 mbar.
1415 LD = 2,400 pbarL/s. Main bath = 0.13 mbar.
1420 LD = 1,580 pbarL/s. Main bath = 0.08 mbar.
1420 Try squirting He around top of fridge, stripline tube and plumbing behind ebox. No noticeable increase in leak signal.
1425 LD = 1,080 pbarL/s. Main bath = 0.05 mbar.
Close valve to scroll pump, let main bath sit at vacuum.
1430 LD = 991 pbarL/s. Main bath = 0.24 mbar.
1430 Try squirting He around top of fridge again. Leak signal does increase a little, 997 to 1,070 pbarL/s, then back down to 1,030.
1435 Let one tube-length of He into main bath, pressure increase from 0.29 to 0.9 mbar. LD 1,030 → 2,350 pbarL/s and climbing. Levels out around 2,700
1440 Let one more tube-length of He into main bath, pressure increase from 1.0 to 2.2 mbar. LD 2,730 → 3,400 pbarL/s and climbing. Levels out around 4,400
1445 Pump on main bath again with scroll pump. LD falling, around 10 pbarL/s per second
1450 LD = 2,600 pbarL/s. Main bath = 0.19 mbar.
1455 LD = 1,360 pbarL/s. Main bath = 0.08 mbar.
1455 Fill main bath with N2. Leak signal falling.
1500 LD = 590 pbarL/s. Main bath = 19/30 bar (mechanical gauge).
1505 LD = 470 pbarL/s. Main bath = 122 mbar over 1 atm (G3).
1510 Try squirting controlled amounts of He down the LN2 fill tube. Very little change in leak signal, maybe 495→500 pbarL/s
1510 Squirt short blast of He down LN2 fill tube, enough to hear gas come out of the other (uncorked) port. Leak signal 560 → 1500 pbarL/s and climbing… rises above 5,000, starting to plateau.
1515 LD = 7,000 pbarL/s
1520 Try a blast of He down the third fridge port, with the recovery port open (ping pong ball cover). In this case the helium will mostly just come in the top and back out the top. Leak signal does not rise. In fact it has been falling the last few minutes.
1523 Another blast of He down the LN2 fill tube. Leak signal starts at 5460 pbarL/s, several second delay, then quick rise. Reaches 7700 in after a couple of minutes.
1530 Raise the LN2 tube 1 inch so it is no longer seated in funnel, then squirt a blast of He down it. Leak signal began at around 8000 pbarL/s, does not rise, in fact it is falling.
1535 Push the LN2 tube back into the funnel, then squirt another blast of He down it. Leak signal began at around 6980 pbarL/s, after a few seconds starts rising rapidly, to 10,000 in a couple minutes.
This strongly suggests that the leak is in the lower part of the IVC, somewhere below the funnel for the LN2 fill.

As shown in the diagram above the funnel is above three seals: the bottom IVC seal, the middle IVC seal, and the stripline tube / IVC seal. So any of those could in principle be where the leak is. But during blowout the leak signal began its steep rise only at the very end of the transfer, so I still suspect the bottom seal.

This is twice that it failed. Both times I (Matt) assembled the indium o-ring. It's possible I've gotten sloppy with the vacuum grease and put too much on. If the seal depends on a significant amount of grease it's not surprising that it would fail when cold.

1630 Begin cryocleaning mixture through Trap 2, V6=12.4%
1635 MC(Cernox) = 77.1 K, Sorb = 132.8 K
1635 Let 2.2 mbar N2 into IVC. This is two tube-lengths. Pressure dropped after closing valve, close to boiling point so pressure may be lower than otherwise. Should be enough, don't want to overdo it.
1640 Increase V6 → 12.8% to reestablish flow
1645 MC(Cernox) = 77.45 K, Sorb = 134.9 K, Cryostat outer surface = 23.3 C
1650 Close IVC valves then move pump line to cryostat vacuum jacket valve
1655 Let 100 mbar N2 gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
1700 MC(Cernox) = 77.7 K, Sorb = 136.6 K, Cryostat outer surface = 23.0 C
1700 Let 100 mbar N2 gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
1700 Increase V6 → 12.9% to reestablish flow
1705 MC(Cernox) = 77.85 K, Sorb = 137.5 K, Cryostat outer surface = 22.5 C
1715 MC(Cernox) = 78.1 K, Sorb = 139.6 K, Cryostat outer surface = 21.9 C
1720 Finish cryocleaning. Pump mixture back to dump.
1730 Open mixture to cold LN2 Trap 1 by opening v9 and v13a
1735 MC(Cernox) = 78.8 K, Sorb = 145.3 K, Cryostat outer surface = 21.0 C
1950 MC(Cernox) = 92.9 K
2110 MC(Cernox) = 105.44 K
warming time constant looks like it's around 21 hours, won't hit 273 K till Saturday evening
Cryocleaning log
date trap duration (minutes) contaminants removed (mbar)
Mar 13 2 70 8.1
Mar 13/14 (overnight, no circulation) 1 overnight 7.5
Mar 14 2 70 8.9
Mar 14 1 70 6.6
Mar 14 2 70 4.8
Mar 14/15 (overnight, no circulation) 1 overnight 10.0
Mar 15 2 50
Mar 15/16 (overnight, no circulation) 1 overnight

Next Day