2018/03/09 (Friday)

Previous Day

Status: Ending run, letting main bath empty.
Overnight we took bias sweep data at ~4.46 K
0930 Tower = 4.5 K, 1K Pot = 6.0 K, Fridge (G1) = 550 mbar. IVC pumping pressure = 5.4 nbar, Main bath = 3.4%.
0945 Begin pumping mixture out of fridge & Cold Trap 1 to dump.
causes a cooling effect on 1KPot/Tower, → 4.9/4.1 K, then start rising again.
Dump (G2) = 731.4 mbar when pumpout finished.
1010 One of the square wave series from yesterday had all noise triggers. Set up square wave injection scheme as yesterday by adding switchboard between DCRC3 and readout cables.
1016 Start 5 min of square wave calibration data with 10Hz, 5mVpp square wave signal on bias electrode, DIB1/2 = DCRC50/55, triggering on DCRC1QI = 300 ADC bins, DCRC bias = 0V, 1000 BOR randoms.
1126 Try to change to other calibration setup from yesterday: DIB1/2 = DCRC33/50. Keep getting jagged noise.
1130 Start up with just DIB1=DCRC33. Noise looks good.
1131 Take 1000 randoms of DIB1=DCRC33 noise data.
1135 Power cycle a few times, now the noise looks fine.
1136 Start 5 min of square wave calibration data with 10Hz, 5mVpp square wave signal on bias electrode, DIB1/2 = DCRC33/50, triggering on DCRC2QI = 120 ADC bins, DCRC bias = 0V, 1000 BOR randoms.
1144 Stop pumping on IVC. IVC = 54 <m>\mu</m>bar. Watching DCRC3QI noise as things warm up. DCRC3QI=0.013Vpp, tower=4.41 K
1150 IVC = 74 nbar, DCRC3QI=0.013Vpp, tower=4.39 K
1200 IVC = 100 nbar, DCRC3QI=0.013Vpp, tower=4.36 K
1235 IVC = 190 nbar, DCRC3QI=0.026Vpp, tower=4.31 K, 1K Pot = 4.98 K.
All four channels are still alive, not railed.
1246 Power down DCRCs.
We'll watch the detector resistance rise during warming.
1248 MC (Cernox)=4.23 K, DIB1 qIbias-QET D = 1.22 nS
1300 IVC = 300 nbar, Tower=4.23 K, 1K Pot = 4.93 K
1300 Disconnect pump manifold from IVC and connect it instead to the cryostat vacuum jacket. Main bath < 0%
1310 Pump out line leading to cryostat vacuum jacket down to 40 ubar (not very low for some reason). Open vacuum jacket valve for a minute - pressure goes down to 10 ubar. So we let a little air in - didn't seem to cause a boiloff.
1315 Let 140 ubar He gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
Looks like this wasn't enough to start a boiloff.
1320 Let 740 ubar He gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
This also wasn't enough.
1325 Actually it looks like maybe this was just enough to cause some warming. Tower temperature is rising slowly. Recovery meter is spinning faster, 1 sqft/24 seconds. Surface of main bath is 21.7 C, about 1 C below room temperature.
1330 1K Pot temperature increased from 4.90 to 4.94 K, but then fell a bit. Probably the helium in the vacuum jacket condensed and is not quite making a thermal connection to the room.
1KPot temperature came up again. Recovery meter still going but slower, 1 sqft/40 seconds.
1335 1KPot/Tower = 4.99/4.36 K, Main bath outer surface = 21.6 C, recovery = 1 sqft/40 seconds. DIB1 qIbias-QET D = 1.84 nS
1350 1KPot/Tower/Cernox = 5.07/4.42/4.45 K, Main bath outer surface = 21.4 C, recovery = 1 sqft/64 seconds. DIB1 qIbias-QET D = 2.33 nS
1400 1KPot/Tower/Cernox = 5.13/4.44/4.48 K, Main bath outer surface = 21.4 C, recovery = 1 sqft/72 seconds. DIB1 qIbias-QET D = 2.57 nS
1410 1KPot/Tower/Cernox = 5.20/4.46/4.51 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/88 seconds. DIB1 qIbias-QET D = 2.93 nS
1420 1KPot/Tower/Cernox = 5.24/4.48/4.53 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/72 seconds. DIB1 qIbias-QET D = 3.12 nS
1440 1KPot/Tower/Cernox = 5.36/4.58/4.64 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/120 seconds. DIB1 qIbias-QET D = 4.39 nS
1500 1KPot/Tower/Cernox = 5.51/4.66/4.72 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/120 seconds. DIB1 qIbias-QET D = 5.74 nS
1520 1KPot/Tower/Cernox = 5.63/4.72/4.80 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/100 seconds. DIB1 qIbias-QET D = 6.88 nS
1540 1KPot/Tower/Cernox = 5.83/4.83/4.92 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/104 seconds. DIB1 qIbias-QET D = 8.81 nS
1600 1KPot/Tower/Cernox = 6.05/4.96/5.04 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/80 seconds. DIB1 qIbias-QET D = 11.15 nS
1625 1KPot/Tower/Cernox = 6.29/5.12/5.19 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/120 seconds. DIB1 qIbias-QET D = 13.90 nS
1700 1KPot/Tower/Cernox = 6.63/5.36/5.47 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/132 seconds. DIB1 qIbias-QET D = 18.56 nS
1720 1KPot/Tower/Cernox = 6.90/5.64/5.71 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/124 seconds. DIB1 qIbias-QET D = 21.67 nS
1755 1KPot/Tower/Cernox = >7/6.18/6.30 K, Main bath outer surface = 21.3 C, recovery = 1 sqft/152 seconds. DIB1 qIbias-QET D = 24.15 nS
1810 Let 740 ubar He gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
Maybe a small increase in recovery rate, 1 sqft/96 seconds
1812 Let 2.5 mbar He gas into vacuum manifold. Then open cryostat vacuum jacket valve to let this into the vacuum jacket.
Maybe a small increase in recovery rate, 1 sqft/96 seconds
1815 1KPot/Tower/Cernox = >7/6.56/6.68 K, Main bath outer surface = 21.1 C, recovery = 1 sqft/32 seconds. DIB1 qIbias-QET D = 24.52 nS
1820 warmup has accelerated
1825 Sorb/Tower/Cernox = 12.2/7.40/7.40 K, Main bath outer surface = 20.8 C, recovery = 1 sqft/34 seconds. DIB1 qIbias-QET D = 24.76 nS
1840 Sorb/Cernox = 15.3/9.03 K, Main bath outer surface = 20.5 C. DIB1 qIbias-QET D = 24.87 nS

Next Day