2018/03/03 (Saturday)

Previous Day

Status: Warmed up after a blockage, also removed and reinstalled tower. Pumping IVC overnight. Helium in cryostat vacuum jacket from warmup.
0750 IVC pumping pressure = 40 nbar, IVC = 1.6 ubar with pump valved out
Cryocleaning log
date trap duration (minutes) contaminants removed (mbar)
Feb 27 2 55 13.5
Feb 27 1 55 0.2
Feb 28 2 140 4.5
try to cool down, discover blockage
Mar 1 2 75 6.7
Mar 2 2 115 4.6
Mar 2 1 75 10.0
Mar 3 2 83 2.8
Mar 3 1 90 5.5
0800 Put LN2 cold traps in bath and fill bath
0810 Close IVC valves at top of fridge, then connect pumping hose to cryostat vacuum jacket.
0815 Start pumping cryostat vacuum jacket with scroll pump. Had 4.6 mbar in it before pumping.
0820 Start cryocleaning mixture through LN2 Trap 2, V6 = 12.5%
0830 Switch to turbo pump to pump out cryostat vacuum jacket
0830 Increase V6 → 12.9% to reestablish flow
0835 Cryostat vacuum jacket pumping pressure = 0.86 ubar
1003 Stop cleaning through trap 2 and pump mixture to dump.
1004 Cryostat vacuum jacket pumping pressure = 0.19 ubar
1038 Remove LN2 Cold Trap 2 from bath to let warm up
1045 Start cryocleaning mixture through LN2 Trap 1, V6 = 13.0%
1057 Increase V6 → 13.2% to reestablish flow
1215 Stop cleaning through trap 1 and pump mixture to dump.
1230 Remove Trap 1 from LN2 bath to let warm up
1235 Check warm Trap 2: around 2.8 mbar. Pump this out with scroll pump.
1250 Cryostat vacuum jacket pumping pressure = 0.12 ubar. Pressure with pump valve out = 14 ubar
1310 Noticed that the LN2 Dewar is low. Took it to the fill station but have to wait for another Dewar being filled right now.
1315 Trap 1 is mostly warmed up (put it in a bucket of hot water to warm it faster). Remove from bucket and check the pressure: 7.8 mbar, but it's falling now, maybe it was a bit warmer than room temp, not sure - should wait for it to stabilize
1355 Trap 1 feels like room temperature, now the pressure reads 5.5 mbar. Pump out trap with scroll pump.
1405 Check pressure in 1K Pot: 28 mbar on P2. Pump out with scroll pump.
1420 LN2 Dewar is now full
1430 Check fridge pressures: about 0.025 mbar in LHe cold trap. No noticeable amount in Still or Condenser. Pump on fridge a few minutes with scroll pump.
1435 Cryostat vacuum jacket pumping pressure = 0.10 ubar. Pressure with pump valve out = 12 ubar
1440 Open fridge lines up to (cold) Trap 1.
1440 Let mixture from pump into fridge through Trap 1, leaving most mixture in the dump. 3He pump off, V13a, V12a+speedi, V3, V1, condenser & still valves all open
1450 Stop pumping on cryostat jacket. Reconnect pumping manifold to IVC pumping line.
1500 Pump out main bath
1500 Connect Leak Detector to IVC. Background = 14 pbarL/s.
1505 Pressurize main bath with He to 255 mbar above 1atm on G3.
LD signal goes up slightly, to 30 pbarL/s. Tolerable.
1510 Let 5.4 mbar He exchange gas into IVC
1515 Begin LN2 transfer
1515 MC(Cernox) = 295.0 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 218 mbar, Level resistor = 1.030 kΩ.
1525 MC(Cernox) = 293.0 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 217 mbar, Level resistor = 1.084 kΩ.
1530 Mini-throughput check: close off Trap 1 and V3. Only V1, condenser speedivalve, still gate valve, and V9 open. Turn on 3He pump, then open V6 to 12.8%. This pumps mix out of the still side and we can see if the pressure goes down on the condenser side. G1 (condenser) started out at 217 mbar, then started falling at about 1.0 mbar/s.
Restore GHS to previous state.
1535 MC(Cernox) = 280.5 K, Sorb = >249 K (max), IVC = 4.0 mbar, G1 (fridge) = 217 mbar, Level resistor = 1.153 kΩ.
1545 MC(Cernox) = 264.3 K, Sorb = >249 K (max), IVC = 3.4 mbar, G1 (fridge) = 216 mbar, Level resistor = 1.208 kΩ.
1550 Another mini-throughput check as before. G1 (condenser) starts out at 215 mbar, then falls at about 2.0 mbar/s. Good sign. We'll keep our fingers crossed.
Restore GHS to previous state: most mixture in dump, some in fridge, both sides open to Cold Trap 1.
1600 MC(Cernox) = 222.6 K, Sorb = >249 K (max), IVC = 2.5 mbar, G1 (fridge) = 215 mbar, Level resistor = 1.339 kΩ.
1615 MC(Cernox) = 185.0 K, Sorb = 215.3 K, IVC = 2.2 mbar, G1 (fridge) = 213 mbar, Level resistor = 1.339 kΩ.
1630 MC(Cernox) = 162.5 K, Sorb = 174.0 K, IVC = 2.2 mbar, G1 (fridge) = 211.5 mbar, Level resistor = 1.339 kΩ.
1634 MC(Cernox) = 158.0 K, Sorb = 166.9 K, IVC = 2.2 mbar, G1 (fridge) = 214 mbar, Level resistor = 1.339 kΩ.
1634 Stop LN2 transfer

Next Day