start › cdms › k100 › run_diary › 20171225

2017/12/25 (Monday)

Previous Day

Setup: DIB1/3/4 → SL 27/19/20 → 3.2 MHz filters → cables 2,1,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3). DCRC4 → HV power supply.
1755 Tower = 21.7 mK
1757 Begin LHe transfer @ MB = 25.9%
1805 Fill LN2 cold trap bath
1852 End LHe transfer @ MB = 91.5%
Another long sequence like a yesterday
1900 Start sequence of 12 30-minute background data series at HV = 70V, Driver Gains = 2x, QET biases = 480/360/440/480/380/490 uA, Front-End Gains = 10/5 on DCRC1/3, 1000 initial randoms, trigger on DCRC1pA threshold = 8, flashing then waiting 30min after each series, using modified triggerfe, using sequencer script DataLoop_70V_30min_groundflash_30min

Next Day