2017/11/02 (Thursday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/3/2 → gender changers → rerouting boards 1/3/4 → DCRC33(2)/DCRC55(1)/DCRC50(3)
1011 Tower/MC/Coldplate = 82.5/57.7/365 mK, Still = 6 mW / 1.24 K / 0.20 mbar, G1/G2 = 299/283 mbar, 1K Pot = 1.649 K / 38 mbar. MB = 22.3%
1016 Fill LN2 cold trap bath.
1017 Start series of data with Vbias(external) = 10.00 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1035 Dump sizes start to fall around dump 7 and reach a minimum at dump 15.
1040 Stop run, ground detector and flash for 1 second with electrode IR LEDs (DCRC1) with with 2 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 25,000 us. Tower temp rises from 82.2 mK to 84.1 mK, falls back to baseline within 3 min.
1045 Start series of data with Vbias(external) = 10.00 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1105 Dump sizes start to fall around dump 7 and reach a minimum at dump 15.
1111 Stop run, ground detector and flash for 1 second with electrode IR LEDs (DCRC1) with with 2 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 5,000 us. Tower temp rises from 825.4 to 93.8 mK and returns to base in 10 min.
1116 Start series of data with Vbias(external) = 10.00 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1130 Dump sizes start to fall around dump 7 and reach a minimum by dump 12.
1158 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 1,000,000 us. Tower temp does not change from 82.7 mK
1203 Start series of data with Vbias(external) = 10.00 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1230 Dump sizes start falling at Dump 4 and reach a minimum by Dump 10.
1241 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 500,000 us. Small tower temp spike from 82.6 to 83.0 mK and back.
1246 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1330 Dump sizes start falling at Dump 4 and reach a minimum by Dump 10.
1337 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 100,000 us. Small tower temp spike from 83.0 to 84.5 mK and back.
1342 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1400 Dump sizes start falling at Dump 3 and reach a minimum by Dump 10.
1414 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 25,000 us. Small tower temp spike from 82.9 to 87.4 mK.
1418 Tower at 83.2 mK and cooling.
1419 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1445 Dump sizes start falling at Dump 4 and reach a minimum by Dump 10.
1451 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 5,000 us. Tower spikes from 82.9 to 103 mK.
1455 Tower at 90 mK and cooling.
1456 Start series of data with Vbias(external) = 10.02 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1515 Dump sizes start falling at Dump 4 and reach a minimum by Dump 9.
1517 Stop run, ground detector and flash for 1 second with electrode UV LEDs (DCRC1) with with -4 mA LEDcurrent, LEDPulseWidth at standard 100 us, and with LEDRepRate at standard 5,000 us. Tower spikes from 83.5 to 117 mK. (Base temperature seemed to be still elevated since the last flash.)
1521 Tower at 95 mK and cooling.
1522 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1515 Dump sizes start falling at Dump 4 and reach a minimum by Dump 9.
we'll try a known good flashing again just to doublecheck
1554 Stop run, ground detector and flash for 1 second with electrode IR LEDs (DCRC1) with with 2 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 25,000 us. Tower temp spikes from 83.0 mK to 87.5 mK. OOPS - no, I only set LED1 to 2 mA, LED2 was at -4 mA. So this was a “hybrid”, flashing with one IR and one UV LED. That's probably why the temperature went higher than last time I tried this configuration, I think the UV heats the tower more.
1558 Tower at 84.2 mK and cooling.
1559 Start series of data with Vbias(external) = 10.01 V, with 1000 beginning randoms, and a 20 ADC bin threshold trigger on DCRC2qO, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1620 Dump sizes start falling at Dump 7 and reach a minimum by Dump 13.
1622 Stop run and ground electrode.
1625 Set up to take another data series while observing the LED pulses. On DCRC1, LEDpulseWidth 10 us, LEDrepRate 100,000 us, connect LED outputs to phonon channels. Turn off randoms, turn off DCRC2 qO triggering, turn on DCRC1 pA & pB triggering. Increase phonon trace to 512 samples.
1630 Start series of data with Vbias(external) = 0 V, with no beginning randoms, and a 100 ADC bin threshold trigger on DCRC1pA,pB, with trigger baseline = 32 samples and trigger length = 10 samples (these are 2x the usual values).
1630 Flash for 1 second with electrode IR LEDs (DCRC1) with with 2 mA LEDcurrent, LEDPulseWidth reduced from 100 us to 10 us, and also with LEDRepRate reduced to 100,000 us.
1631 Stop data.
1635 Looking at data with BatViewer, looks like 19 LED pulses. 19*100,000 us = 1.9 seconds. So yes it looks like there's about a 1 second lag in the sequencer.
1635 Close V6 to stop circulation and allow mix to expand into still line
1635 Turn off still heater
1640 Fill LN2 cold trap bath (will pump out trap tomorrow)
1640 Tower temperature increasing, Still pressure increasing, Condenser pressure falling
1640 IVC = 30 nbar. We have not been pumping on it. A little higher than last run
1645 Tower = 280 mK, Still = 5.5 mbar, Condenser = 185 mbar.
1645 Close 1K Pot speedivalve to stop pumping on pot. Leaving needle valve open. Close V4a and valve to 4He pump. Open 4He outlet to “exhaust” and close recovery valve. Turn off 4He pump.
1650 Turn off 3He pump. Since v6 and v9 are closed, pump is isolated.
1650 Tower = 506 mK, Still = 140 mbar, Condenser = 118 mbar. 1K Pot = 2.32 K / 0.018 mbar.
1655 Open V4 to let condenser expand into dump. Condenser falls to 20 mbar.
1700 Open V3 to let still side expand into dump. Condenser falls slightly.
1700 Main bath = 11.4%
1700 Tower = 617 mK. Leaving the tower thermometer on with range reduced to 6.32 kΩ and excitation 20 uV
1710 Tower = 667 mK, Fridge = 22 mbar. 1K Pot = 3.12 K
Date Time of
flash
LEDs used Flash
time [s]
Time between
flashes [us]
Pulse
width [us]
Max tower
temp [mK]
Wait time
[min]
Dump at which
sizes start falling
Dump at which
minimum reached
Oct-31 11:24 DIB IR 60 5000 100 700 8 13
Oct-31 12:20 DIB IR 6 5000 100 360 7 13
Oct-31 12:54 DIB IR 2 5000 100 255 7 13
Oct-31 14:56 DIB IR 2 5000 100 240 7 12
Oct-31 16:11 DIB IR 2 5000 10 135 7 12
Oct-31 16:55 none 6 4 9
Nov-1 09:05 none o/n 7 13
Nov-1 09:41 none 25 7 13
Nov-1 11:00 none 5 4 7
Nov-1 12:05 DIB IR 2 5000 100 225 7 12
Nov-1 12:33 DIB IR 2 5000 10 123 7 12
Nov-1 13:04 none 5 3 8
Nov-1 13:35 DIB IR 2 25000 10 90 7 12
Nov-1 14:06 DIB IR 2 100000 10 84.4 7 12
Nov-1 14:36 DIB IR 2 500000 10 81.85 6 12
Nov-1 15:04 DIB IR 2 1000000 10 82.1 5 11
Nov-1 15:59 electrode IR 2 1000000 10 5 12
Nov-1 16:33 electrode IR 2 100000 10 82.4 6 12
Nov-2 10:40 electrode IR 2 25000 10 84.1 7 15
Nov-2 11:11 electrode IR 2 5000 10 93.8 7 12
Nov-2 12:41 electrode UV 2 500000 10 83 4 10
Nov-2 13:37 electrode UV 2 100000 10 84.5 3 10
Nov-2 14:14 electrode UV 2 25000 10 87.4 4 10
Nov-2 14:51 electrode UV 2 5000 10 103 4 9
Nov-2 15:17 electrode UV 2 5000 100 117 4 9
Nov-2 15:54 electrode IR/UV 2 25000 10 87.5 7 13