| 1633 | Power cycle and get DCRC1 stuck again. rd 14 = 377F (stuck), but rd 2, rd 3 = updating. It looks like the SDRam address is being updated but the data coming out is not. |
| 1700 | Played a little more with this. The above “good” values are when first powered on, but without programs etc. running. It seems When first turned on, both the address (2,3) and the data(14) are 0000. Once the programs and run are started, 2, 3, and 14 are all updating. When the DCRC gets “stuck”, 2,3 continue to update, but 14 does not. If the run is then stopped, 2,3 stop updating. Looking at the description of these registers, it seems like the DAQ is setting 2,3 to request data and the DCRC fills 14 with data from the requested memory location. |
| 1705 | Maybe writing a bunch to 0x02,0x03 puts it in the stuck state? Wrote a script to write an address to 0x02, 0x03 10,000 times in a row, but DCRC1 continued to work after that. |
| 1720 | Close V6 to stop circulation and allow mix to expand into still line |
| 1720 | Turn off still heater |
| 1730 | Fill LN2 cold trap bath (will pump out trap tomorrow) |
| 1730 | G1/G2 have fallen to 90 mbar, Tower has risen to 330 mK, Still pressure has risen to 8 mbar. |
| 1730 | Close 1K Pot speedivalve to stop pumping on pot. Leaving needle valve open. Close V4a and valve to 4He pump. Open 4He outlet to “exhaust” and close recovery valve. Turn off 4He pump. |
| 1735 | Turn off 3He pump. Since v6 and v9 are closed, pump is isolated. |
| 1740 | Open V4 to let mixture on condenser side and in trap expand into dump. |
| 1740 | IVC = 16 nbar. (We have not been pumping on it this run.) Comparable to last run. During that run we pumped on the IVC continuously. |
| 1740 | Tower = 500 mK |
| 1745 | Tower = 500 mK, G1 (condenser/trap/dump) = 13.5 mbar, G2 (3He pump, off) = 3 mbar, Still = 70 mbar and rising, 1K Pot = 2.56 K and rising. |
| 1745 | Open V3 so still side can expand directly to dump too. Pressures don't change much so I guess still pressure had already equalized mostly with condenser pressure. |
| 1750 | Switch to Cernox thermometer. We're below its temperature range - T.UNDER on LakeShore bridge, about 0.1 K on LabVIEW, which is wrong. Overnight it oughta warm up into range. Set LakeShore excitation to 63.2 uV, range 632 kOhm. |