start › cdms › k100 › run_diary › 20170930

2017/09/30 (Saturday)

Previous Day

Status: stopped circulation yesterday. 3He pump is off, small amount of mixture left in it. Mixture open between dump, Trap 1 (cold), still&condenser - V13a, V12a, V4, V3, V1, condenser & still valves open. 1K Pot speedivalve closed, needle valve open. 4He pump off and isolated. MB was below 20% last night, should be below 0% now (though likely still some in the bottom).
1120 G1/G2 (trap+dump+fridge) = 215/185 mbar, rising steadily overnight. MC(Cernox) = 2.1 K, slow rise overnight. 1K Pot > 7K (max). Sorb = 11.8 K. IVC = 22 nbar.
1125 Close 1K Pot needle valve, isolating pot with He gas inside.
1145 Finish pumping Cold Trap 1 to dump. Remove both traps from bath to let them warm up.
1145 Start pumping mixture out of fridge (both sides) to dump. Might still be too cold to get it all out.
Sorb, Still, MC temperatures fall. So mixture is still condensed. G1 (fridge) fell quickly to 60 mbar, now falling more slowly. G2 (dump) rises to 330 mbar, rising slowly.
1150 V6 12.8→14.0%, to speed up pumpout. Pressures change more quickly but cooling effect also enhanced.
1205 G1/P2 (fridge) = 0.9/0.22 mbar, G2 (dump) = 382 mbar, MC(H35)/Still/Sorb = 1.13/1.54/8.9 K. G2 rising very slowly. Fridge still too cold to pump out efficiently.
1210 Close V6 to stop pumping on fridge. Then turn on MC heater and Still heater to 20 mW each.
1220 G1(fridge, now isolated and being heated) has risen to 40 mbar. MC(H35) has risen to 2K.
1220 Let 1 hose-length of He gas into IVC. Much quicker warmup and increase in fridge pressure.
G1(Fridge) → 400 mbar and rising. MC(Cernox) → 4.6 K. IVC → 34 μbar.
1225 With fridge at 650 mbar and rising, open V6 to 13.0% to start pumping to dump again.
1230 MC(Cernox)/Sorb = 4.7(flat)/8.1(rising) K.
1230 G1(fridge)/G2(dump) = 105(falling)/735(rising) mbar. Open V6 to 14% to speed up pumpout.
1230 Turn off MC and Still heaters, shouldn't need them. MC temperature falls a little.
1235 G1/P1(fridge) = 1.6/0.4 mbar, falling. G2(dump) = 767 mbar, rising very slowly. Open V6 to 25%.
1240 MC(Cernox)/Sorb = 4.64(flat)/8.3(flat) K.
1245 G1/P1(fridge) = 0.9/0.03 mbar, P1 still falling a little. G2(dump) = 767.5 mbar, nearly flat.
1300 G1/P1(fridge) = 1.0/0.02 mbar, P1 essentially flat. G2(dump) = 768 mbar. Isolate fridge, close IGH valves, turn off 3He pump.
1300 MC(Cernox)/Sorb = 4.67(slow rise)/8.5(slow rise) K. Good, we've retrieved the mix with the fridge still below H2 boiling point. This time we'll be able to pump any H2 in the LHe cold trap out. Not to dump of course, pump out to air.
1310 Check fridge port - definitely significant pressure, still LHe down there.
1315 Let 3-inch hose length of He gas into cryostat vacuum jacket. Recovery meter starts turning swiftly. MC and Sorb temperatures rise.
1315 He recovery = 1 cuft / 6 seconds
1320 He recovery = 1 cuft / 7 seconds. So it's not getting faster, probably OK to leave it like this. Recovery hose a little frosty, Cu warming coil only a little frosty.
1320 MC(Cernox)/Sorb = 6.10/14.1 K, both rising
1330 MC(Cernox)/Sorb = 8.34/18.4 K, both rising. He recovery = 1 cuft / 11 seconds. Cryostat somewhat cool to the touch. IVC = 63 μbar (up from 34)
Probably could stand to put extra gas in the cryostat jacket, but I'll leave it for now. Maybe tomorrow (or Monday) it will be warm enough to put N2 in instead of He. Easier to pump out N2 and not as bad if you don't get it all out.
1335 Start new temperature logging file, I:\K100\temperature_20170930.txt
1925 MC(Cernox) = 98.7 K

Next Day