UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20170802

2017/08/02 (Wednesday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/2/3 → gender changer 1/3/4 → DCRC55(1)/33(2)/X(4)
0909 Tower = 25.7 mK, Still = 0.060 mbar, G1/G2 = 146/135 mbar, 1K Pot = 1.69 K / 30 mbar. MB = 65.5%
0910 Fill LN2 cold trap full.
0952 Start takeAndFlash sequence of 4 60-minute data series of Cf data with Frontend Gain = (15/2)/(4/4)/(5/5/5/5), OG = 4/2/4, QET biases = 2V setting from 2017/07/25, with 1000 beginning randoms, 30 sec flash with 30 min wait, and a 15ADC bin threshold trigger on DCRC4pB (Side2pF). Charge biases are DCRC1/2 (Side2/1) = -2/2 V.
1030 Sequence aborted remotely to do DAQ work
1235 DAQ work is finished. Detector still biased. Set charge bias to 0 V then flash for 30 seconds.
1401 Start takeAndFlash sequence of 2 60-minute data series of Cf data with Frontend Gain = (15/2)/(4/4)/(5/5/5/5), OG = 4/2/4, QET biases = 2V setting from 2017/07/25, with 1000 beginning randoms, 30 sec flash with 30 min wait, and a 15ADC bin threshold trigger on DCRC4pB (Side2pF). Charge biases are DCRC1/2 (Side2/1) = -2/2 V.
1725 Fill LN2 cold trap bath

Next Day