2017/07/28 (Friday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/2/3 → gender changer 1/3/4 → DCRC55(1)/33(2)/X(4)
0831 Tower = 25.6 mK, Still = 0.060 mbar, G1/G2 = 159/146 mbar, 1K Pot = 1.70 K / 32 mbar. MB = 34.8%
0833 Fill LN2 cold trap full.
1004 Tower/MC/Coldplate = 25.9/28.5/171 mK, Still = 1.3-1.5 K / 0.060 mbar, G1/G2 = 160/146 mbar, 1K Pot = 1.70 K / 34 mbar. MB = 32.3% (predicted level: 32.5%)
1034 Start LHe transfer (MB = 31.8%)
1117 End LHe transfer (MB = 94.2%)
1130 Release pressure from MB.
1139 Start 10 min series of 20Hz Randoms, Tower = 25.6±0.1 mK
1153 Apply 0.5 uW to mixing chamber
1206 Start 10 min series of 20Hz Randoms, Tower = 26.7±0.1 mK
1222 Apply 1.5 uW to mixing chamber
1228 Start 10 min series of 20Hz Randoms, Tower = 36.0±1.0 mK
Also start a temperature logfile on vuk-01: “I:\K100\temperature_20170728_1p5.txt”
1247 Apply 2.5 uW to mixing chamber
1255 Start 10 min series of 20Hz Randoms, Tower = 42.8±2.3 mK
1309 Apply 3.5 uW to mixing chamber
1312 Start 10 min series of 20Hz Randoms, Tower = 46.5±1.5 mK
1323 Turn of MC heater
1400 Tower = 25.8 mK
1401 Set DCRC1 phonon trace length = 10 ms
1403 Start frequency scan sequence with 6mVpp signal on DCRC1 PA,PC, files temporarily stored as /livedata/Fscan_[frequency].mid.gx Problem with sequence. Data deleted
1449 Start frequency scan sequence with 6mVpp signal on DCRC1 PA,PC, files temporarily stored as /livedata/Fscan_[frequency].mid.gx
Created a mock series: “midasraw/byseries/07170728_2500” and store this scan data there
1503 Start frequency scan sequence with 12mVpp signal on DCRC1 PA,PC, files temporarily stored as /livedata/Fscan_[frequency].mid.gx
Created a mock series: “midasraw/byseries/07170728_2600” and store this scan data there
1522 Restore DCRC1 phonon trace lengths to 3.2768 ms
1536 Looking at DCRC1PA with 100ms trace length, estimate rate to be 112 Hz.
1640 Fill LN2 cold trap full.
1652 Start sequence 12 60-minute data series of charge-bias-scan data with Frontend Gain = (15/2)/(4/4)/(5/5/5/5), OG = 4/2/4, QET biases = 2V setting from 2017/07/25, with 1000 beginning randoms, 30 sec flash with 30 min wait, and a 15ADC bin threshold trigger on DCRC4pB (Side2pF). Charge biases are DCRC1/2 (Side2/1) = -/+. Side 2 biases in order are 0, -0.25, -0.5, -0.75, -1, -1.5, -2, -2.5, -3, -3.5, -4, -5