Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
0910 Tower = 33.1 mK Main Bath = 30.6%
1230 Refill LN2 dewar.
1239 Tower/MC/Coldplate = 33.0/29.6/191 mK, Still = 1.26 K / 0.110 mbar, G1/G2 = 222/207 mbar, 1K Pot = 1.674 K / 22.0 mbar. Main Bath = 25.6% (predicted level: 25.8%)
1248 Start LHe transfer (MB = 25.2%)
1341 End LHe transfer (MB = 91.3%)
1400 Fill LN2 cold trap bath.
1410 Created an IbIs sequence, IbIs_sequence3.msl, which steps from 250 uA down to 50 uA in 0.2 uA steps, taking 5 seconds of data at each step as a separate MIDAS run. Need to disable Write As CDMS Series in /Logger, and need to remember to reenable this.
1440 Set all Driver Gains to 1. Set QET biases to the IbIs peak values (140/155/140/-135/-160/-155). Then adjust Driver Offsets to set baseline high, around 1.1 V. Driver offsets: -0.6/0.0/-0.6/0.11/0.20/0.12. This should give enough Vout range to capture entire IbIs curve.
1451 Start IbIs sequence (start #6846) in 1-sided 0 V mode with Frontend Gain = (15/10/10)/(5/4/4), OG = 1/1, with QET biases starting at 250/-250 stepping down in 0.2 steps to 50/-50, single 5 second MIDAS run per step, 20 Hz randoms.
1500 Tried copying the first file as a pseudo-CDMS file in directory /byseries/07170602_2500. But couldn't read it with BatViewer. Stop sequence to test whether an ordinary run will read properly.
1506 Start test data series with settings as they were left off in the sequence but with Write as CDMS Series enabled. Just hit the start button then stop after a few seconds.
1510 This was also unreadable. Try restarting the DAQ programs.
1510 Start test data series with settings as they were left off in the sequence but with Write as CDMS Series enabled. Just hit the start button then stop after a few seconds.
1515 This was also unreadable - using jiajin. But after ssh-ing into cdms1, BatViewer works on all the data taken today. Ok then.
1518 Start IbIs sequence (start #6893) in 1-sided 0 V mode with Frontend Gain = (15/10/10)/(5/4/4), OG = 1/1, with QET biases starting at 250/-250 stepping down in 0.2 steps to 50/-50, single 5 second MIDAS run per step, 20 Hz randoms.
1652 Sequence has reached 186.2 uA. At this rate it will finish by 20:12.
1847 It looks like the DAQ failed somehow at this point. The last good run is #7615 with the QET biases set to 105.6. Then the next run continued throughout the weekend with one dump per minute and very small files. Deleted all that data.
The first set of IbIs data, started at 14:51 (Midas start #6846) has been copied to /byseries/07170602_2500.
The second set of IbIs data, started at 15:18 (Midas start #6893) has been copied to /byseries/07170602_2600.
In both directories the dump number indicates the QET bias times 10, e.g. 07170602_2600_F1714.mid.gz is with QET bias = 171.4 uA.